5. TECHNICAL SPECIFICATION
5.1. Functions
Continuity
Operating range of use according to EN 61557-4: 0,028 Ω ÷ 200 Ω.
Test current ≥ 200 mA at R ≤ 2 Ω.
Measuring
range (Ω)
Resolution (Ω)
Reference error
Operating error
0,000 ÷ 4,000
0.001
±
(3 %
z MH + 5 D)
±
(4 %
z MH + 7 D)
4,00 ÷ 20,00
0.01
±
(2 %
z MH + 5 D)
±
(3 %
z MH + 7 D)
20,0 ÷ 200,0
0,1
±
(2 %
z MH + 4 D)
±
(3 %
z MH + 6 D)
DC voltage
Measuring range
(V)
Resolution (V)
Reference error
Operating error
0,000 ÷ 2,000
0.001
±
(3 %
z MH + 10 D)
±
(4 %
z MH + 12 D)
2,00 ÷ 43,00
0.01
±
(2 %
z MH + 5 D)
±
(3 %
z MH + 7 D)
43,0 ÷ 650,0
0.1
±
(2 %
z MH + 5 D)
±
(3 %
z MH + 7 D)
AC voltage
(TRMS value)
Measuring range
(V)
Resolution (V)
Reference error
Operating error
0,000 ÷ 2,000
0.001
±
(3 %
z MH + 10 D)
±
(4 %
z MH + 12 D)
2,00 ÷ 30,00
0.01
±
(2 %
z MH + 6 D)
±
(3 %
z MH + 8 D)
30,0 ÷ 460,0
0.1
±
(2 %
z MH + 6 D)
±
(3 %
z MH + 8 D)
Specified accuracy is valid for THD < 2 %
Frequency range 48 ÷ 52 Hz.
Notes to the parameters stated in chapter 5.1:
a) Stated accuracy of Continuity measurement is valid only if no parallel circuits
are connected to measured resistance.
b) Thresholds of auto-ranging may differ from the stated values.
c) R... Reading, D... Digit.