DIGIOHMpro

- 10 -

Manual range selection

After entering the R function measuring range is chosen automatically.

If necessary, you can press and release the

RANGE

key to select a certain

range, thus disable and eventually re-enable auto-ranging.

Important notices for resistance measurement

-

Do not connect test tips to external voltage

in the R function. Such status is

indicated by beeper and the symbol "Uin", but despite it the instrument can be

damaged!

- Measurement of low resistance with 200 mA current heavily loads the supply

accumulators /batteries of the instrument.

Carry out resistance measurement

only for necessary period of time

. If you see a low battery indicator or if there

are other problems, for example spontaneous shutdown of the instrument,

display flickers, malfunction of control keys etc., insert new fully charged NiMH

batteries, or the quality alkaline batteries. See chapter 4.1 for instructions how to

do it.

3.3.3. Function "●›)" - beeper

The „●›)“ function is intended for measurement of resistance in range 0 - 4 Ω. If

measured value is between 0 - 4 Ω, it is signalized by beeper. Follow the same

measuring procedure and consider the same warnings that are stated for the

resistance function R (see chapter 3.3.2.).

3.4. Phase (live) conductor test, phase sequence test, DIGISORT

Repeatedly press and release the

FUNC

key to select between the following

measurement functions:

3.4.1. Single-pole / contactless phase indication

This function is suitable for contactless indication of the presence of voltage on

conductor / device and contact single-pole indication of the phase conductor.

Press and release the

FUNC

key to select the phase indication function. The

right side of the display briefly displays bar graph "-------------" and the left side

displays "X" on a blue field.

Approaching the test tip on the instrument’s case to the conductor / device

under voltage bar graph indicates the growing intensity of the electric field.

Sensitivity can be switched in two steps by the

RANGE

key.

Acoustic signalization can be switched on / off by the

CAL / ELEC

key

Touching the test tip on the instrument’s case to the measured conductor with

voltage ≥ 190 V / 48 ÷ 52 Hz against the earth, the "L" symbol is displayed on

the left side of the display.

Summary of Contents for DIGIOHMPro

Page 1: ...USER S MANUAL DIGIOHMpro V 1 2 EN...

Page 2: ...and resistance measurement 8 3 3 1 Function U 8 3 3 2 Function R 9 3 3 3 Function beeper 10 3 4 Phase live conductor test phase sequence test DIGISORT 10 3 4 1 Single pole contactless phase indicatio...

Page 3: ...The instrument cables connectors plugs or accessories exhibits visible damages The instrument was stored under unfavourable conditions for a long period The instrument was exposed to extraordinary str...

Page 4: ...ic field such as watches credit cards with magnetic strips etc Images in this manual are illustrative and may vary slightly from the actual state 1 2 General description of the instrument The DIGIOHMp...

Page 5: ...e hand it to a collection point of secondary raw materials in accordance with local regulations Batteries Please dispose of used batteries in the designated locations in accordance with local regulati...

Page 6: ...e safely hidden Against accidental ejection are both parts secured by non contact magnetic latch Use original accessories only Max allowed voltage between test tip and ground is 300V Max allowed volta...

Page 7: ...ched on off by briefly pressing and releasing The U R key It toggles between measurement of voltage resistance and beeper functions The FUNC key It toggles between one pole phase phase rotation and DI...

Page 8: ...st lead black 2 m P 3011 test tip black P 4011 crocodile clip black P 8060 transmitter DIGIsort enables to distinguish deenergized conductors Note optional accessories P 2011 P 3011 respectively P 201...

Page 9: ...ew fully charged NiMH batteries or the quality alkaline batteries See chapter 4 1 for instructions how to do it If the device measures the voltage and if resistance is not measured check the fuse whic...

Page 10: ...out test lead compensation before measurement After test lead resistance compensation is carried out it is automatically subtracted from all resistance measurements and thus only actual value of meas...

Page 11: ...in range 0 4 If measured value is between 0 4 it is signalized by beeper Follow the same measuring procedure and consider the same warnings that are stated for the resistance function R see chapter 3...

Page 12: ...rd phase conductor The display shows the phase sequence 1 2 3 or 1 3 2 and the symbol of a rotating wheel in the direction of the expected phase rotation Displaying is finished after about 10 seconds...

Page 13: ...ltage 3 6 RESET of the instrument If the device does not operate as described in this manual we recommend RESET disconnect both test tips from tested object and turn off the instrument It the proper f...

Page 14: ...ent uses four AAA either alkaline cells or NiMH accumulators The batteries accumulators are continuously monitored see description in the Chapter 3 2 If batteries accumulators are low it must be repla...

Page 15: ...accumulators should return to normal The above described cycle in stand alone intelligent charger is recommended every few months to make If after several cycles of the above described discharge char...

Page 16: ...ukewarm soap water for plastic case cleaning Do not spill cleaning liquid over the instrument Do not use cleaning liquids based on petrol hydrocarbons etc 4 3 Calibration Measuring instruments should...

Page 17: ...1 3 z MH 10 D 4 z MH 12 D 2 00 43 00 0 01 2 z MH 5 D 3 z MH 7 D 43 0 650 0 0 1 2 z MH 5 D 3 z MH 7 D AC voltage TRMS value Measuring range V Resolution V Reference error Operating error 0 000 2 000 0...

Page 18: ...0 C relative humidity max 85 non condensing mains voltage 190 260 V 48 52 Hz instrument s position arbitrary Storage conditions ambient temperature 10 70 C relative humidity max 90 10 40 C non condens...

Reviews: