background image

DIGIOHMpro

- 14 -

4.1.2. Charging of accumulators

For charging of accumulators use only adapter supplied as

optional accessories!

Accumulators are charged as soon as the adapter is connected to mains and to

socket for accumulators charger (see Fig. 2.1). If accumulators are fully

discharged, the charging takes about 6 hours (applies to batteries with a

capacity of 800 mAh). Prolonged charging is not a problem; however, do not

charge accumulators for more than 12 hours.

Notes:

Do not charge alkaline cells – it may lead to explosion, leakage, etc. This can

cause serious damage or destruction of instrument.

During charging of new accumulators or ones that were unused for a longer

period (few months) unpredictable chemical processes may arise. As a result,

the instrument operation time can be significantly reduced. In this case, we

recommend several charge (with optional charger) / discharge (normal use of

the instruments) cycles.

Another way is to use a stand-alone intelligent charger which discharge /

charge each cell individually. The discharge / charge cycle is automatically

executed, see instruction manual for the charger used.

After the procedure, the capacity of the accumulators should return to normal.

The above described cycle in stand-alone intelligent charger is recommended

every few months to make.

If after several cycles of the above described discharge / charge capacity of

the accumulators does not return to normal, this may be due to the fact that

the one or more accumulators are degraded - whereas, the built-in

accumulator charger charges all cells connected in series at the same time,

and even one bad (or just different) cell negatively affects the entire

accumulator pack.

It may result in uneven charging of cells, excessive heating of the cell(s)

during charging etc.

In this case, we recommend that a faulty cell is identified with an intelligent

stand-alone charger, or at least comparing the voltage of each cell and then a

faulty cell replace with a new one.

The above-described effects can not be confused with a normal reduction in

accumulators’ capacity over time. All accumulators with a growing number of

charge / discharge cycles gradually loose capacity. This is normal, depending on

accumulator type, the number and parameters of the discharge / charge cycles.

4.1.3. Replacing the fuse

Replace the fuse by the same type only:

GF F0.5A/500V, breaking capacity 1500A, dimensions 32x6,3mm.

Using of another type of fuse can cause damage of the

instrument and/or operator’s safety can no longer be guaranteed!

Summary of Contents for DIGIOHMPro

Page 1: ...USER S MANUAL DIGIOHMpro V 1 2 EN...

Page 2: ...and resistance measurement 8 3 3 1 Function U 8 3 3 2 Function R 9 3 3 3 Function beeper 10 3 4 Phase live conductor test phase sequence test DIGISORT 10 3 4 1 Single pole contactless phase indicatio...

Page 3: ...The instrument cables connectors plugs or accessories exhibits visible damages The instrument was stored under unfavourable conditions for a long period The instrument was exposed to extraordinary str...

Page 4: ...ic field such as watches credit cards with magnetic strips etc Images in this manual are illustrative and may vary slightly from the actual state 1 2 General description of the instrument The DIGIOHMp...

Page 5: ...e hand it to a collection point of secondary raw materials in accordance with local regulations Batteries Please dispose of used batteries in the designated locations in accordance with local regulati...

Page 6: ...e safely hidden Against accidental ejection are both parts secured by non contact magnetic latch Use original accessories only Max allowed voltage between test tip and ground is 300V Max allowed volta...

Page 7: ...ched on off by briefly pressing and releasing The U R key It toggles between measurement of voltage resistance and beeper functions The FUNC key It toggles between one pole phase phase rotation and DI...

Page 8: ...st lead black 2 m P 3011 test tip black P 4011 crocodile clip black P 8060 transmitter DIGIsort enables to distinguish deenergized conductors Note optional accessories P 2011 P 3011 respectively P 201...

Page 9: ...ew fully charged NiMH batteries or the quality alkaline batteries See chapter 4 1 for instructions how to do it If the device measures the voltage and if resistance is not measured check the fuse whic...

Page 10: ...out test lead compensation before measurement After test lead resistance compensation is carried out it is automatically subtracted from all resistance measurements and thus only actual value of meas...

Page 11: ...in range 0 4 If measured value is between 0 4 it is signalized by beeper Follow the same measuring procedure and consider the same warnings that are stated for the resistance function R see chapter 3...

Page 12: ...rd phase conductor The display shows the phase sequence 1 2 3 or 1 3 2 and the symbol of a rotating wheel in the direction of the expected phase rotation Displaying is finished after about 10 seconds...

Page 13: ...ltage 3 6 RESET of the instrument If the device does not operate as described in this manual we recommend RESET disconnect both test tips from tested object and turn off the instrument It the proper f...

Page 14: ...ent uses four AAA either alkaline cells or NiMH accumulators The batteries accumulators are continuously monitored see description in the Chapter 3 2 If batteries accumulators are low it must be repla...

Page 15: ...accumulators should return to normal The above described cycle in stand alone intelligent charger is recommended every few months to make If after several cycles of the above described discharge char...

Page 16: ...ukewarm soap water for plastic case cleaning Do not spill cleaning liquid over the instrument Do not use cleaning liquids based on petrol hydrocarbons etc 4 3 Calibration Measuring instruments should...

Page 17: ...1 3 z MH 10 D 4 z MH 12 D 2 00 43 00 0 01 2 z MH 5 D 3 z MH 7 D 43 0 650 0 0 1 2 z MH 5 D 3 z MH 7 D AC voltage TRMS value Measuring range V Resolution V Reference error Operating error 0 000 2 000 0...

Page 18: ...0 C relative humidity max 85 non condensing mains voltage 190 260 V 48 52 Hz instrument s position arbitrary Storage conditions ambient temperature 10 70 C relative humidity max 90 10 40 C non condens...

Reviews: