113
8.4.5 System Commands and Companion Queries
These commands are used to modify the system parameters for the instrument.
These commands require a parameter value to be included with the command.
The companion query command will read the parameter using the same value
that is used for setting the parameter.
Command
Description
AC Program Mode
AC Manual
Mode
DC Program Mode
DC Manual
Mode
Unit
PLC X
PLC X
X=0~1, 0=OFF, 1=ON
PLC?
PLC?
0~1
AR X
AUTO RUN X
X=0~1,0=PROGRAM,1=MANUAL
AR?
AUTO RUN?
0~1
OM X
OUT MODE X
X=0~1,0=AC,1=DC
OM?
OUT MODE?
0~1
SS X
SINGLE STEP
X
X=0~1, 0=OFF,1=ON
X=0~1,0=OFF,1=ON
SS?
SINGLE
STEP?
0~1
0~1
ALARM X
ALARM X
X=0~9, 0=OFF,9=high
ALARM?
ALARM?
0~9
CONTRAST
X
CONTRAST X
X=1~9, 9=high
CONTRAST?
CONTRAST?
1~9
PUP X
POWER UP X
X=0-2,0=OFF,1=ON,2=LAST
PUP?
POWER UP?
0-2
LC XXXX
LOOP CYCLE
XXXX
XXXX=0~9999 ,0=Cont,1=OFF
XXXX=0~9999 ,0=Cont,1=OFF
LC?
LOOP
CYCLE?
0-9999
0-9999
VHI XXXX
VOLT HI XXXX
XXXX=5.0~300.0
XXXX=5.0~420.0
V
VHI?
VOLT HI?
5.0~300.0
5.0~420.0
V
VLO XXXX
VOLT LO
XXXX
XXXX=5.0~300.0
XXXX=5.0~420.0
V
VLO?
VOLT LO?
5.0~300.0
5.0~420.0
V
FHI XXXX
FREQ HI
XXXX
XXXX=40.0~1000
Hz
FHI?
FREQ HI?
40.0~1000
Hz
FLO XXXX
FREQ LO
XXXX
XXXX=40.0~1000
Hz
FLO?
FREQ LO?
40.0~1000
Hz
SAG XXXX
START
ANGLE XXXX
XXXX=0-359
°
SAG?
START
ANGLE?
0-359
°
EAG XXXX
END ANGLE
XXXX
XXXX=0-359
°
EAG?
END ANGLE?
0-359
°
RESULTS X
RESULTS X
X=0-2, 0=ALL,1=P/F,2=LAST
°
RESULTS?
RESULTS?
0-2
°