49
042009
User defined solvents
☞
Select one of the options NewSolvent1 ... NewSolvent5 from the
"Solvents" screen.
☞
Select one of these solvents and press the rotary/push knob.
All the parameters displayed for the solvent can now be modified.
☞
Turn the rotary/push knob to the right/left to select the parame-
ter to be changed.
☞
Press the rotary/push knob.
☞
Turn the rotary/push knob to the right/left to change the selected
parameter.
☞
P
ress the rotary/push knob.
☞
Press SET to save all values.
The name of the solvent will be changed to "UserSolvent1". This
action also selects UserSolvent1 as the current solvent.
Note: It is not possible to return to the selection list.
Auto-distillation Mode “ menu
"Auto-distillation Mode” display screen
1. Title
2. Control field for solvent
3. Vacuum display (actual value)
4. Vacuum display (target value)
5. Control field for „Start/ Stop“ distillation
6. Timer display
7. . Distillation progress indicator
(% of required quantity of distillate)
8. Vacuum-rotation-time graph
9. Display of flow rate, differential temperature, and condenser heat
rejection rate.
10. Rotary speed display (target value) and speed measurement unit
11. Error message in case of fault, otherwise status indicator
Changing the selected solvent
☞
Turn the rotary/push knob to the right/left to select the control
field "Solvent".
☞
Press the rotary/push knob to display a list of the various solvents
available.
☞
Turn the rotary/push knob to the right/left to select a "Solvent".
☞
Press the SET key.
The solvent selected will be used for the subsequent distillation.
Note: The display shows the parameters for each of the solvents.
The parameters can only be changed for the user-definable solvents
UserSolvent1 ... UserSolvent5. Please refer to the next section for
instructions on how to do this.
1
11
10
2
3
8
7
6
5
9
4