43
042009
Abbreviations used:
m
=
Numbering parameter (integer)
X
=
2
Temperature heating plate
X
=
3
Safety temperature heating plate
X
=
4 Speed
X
=
60
Interval time (1-99 seconds, 1 <= m >=99)
X
=
61
Timer (1-199 minutes, 1 <= m >=199)
X
=
62
Upper lift position (OUT_SP_62 1-> drive lift up)
X
=
63
Lower lift position (OUT_SP_62 1-> drive lift down)
X
=
66
Value vacuum controller
X
=
70
Vacuum controller hysteresis
X
=
74
Tempering medium (OUT_SP_74 0=Öl, OUT_SP_74 1=Wasser)
NAMUR Commands
Function
IN_NAME
Request designation
IN_PV_X
Read actual value
X = 2, 3, 4, 66
IN_SOFTWARE
Request software Id-number, date and version
IN_SP_X
Read target value input
X = 2, 3, 4, 60, 61, 66, 70
OUT_SP_X m
Set target value to m
X = 2, 3, 4, 60, 61, 62, 63, 66, 70, 74
RESET
Switch to normal operation
START_X
Switch on appliance (remote) function
X = 2, 4, 60, 66
STOP_X
witch off appliance function
X = 2, 4, 60, 66
PC 1.1 Cable
This cable is required to connect the 9-pin connector to a PC. (fig. 24).
Fig. 24