Error codes

Any malfunctions during operation will be identified by an error message on the display�
Proceed as follows in such cases:

-

 Switch off device using the main switch at the back of the device

-

 Carry out corrective measures

-

 Restart device

Error code Effect

Cause

Solution

Err 01

Pump off
Heating/refrigerating off

No external sensor

- Check this sensor

Err 02

Pump off
Heating/refrigerating off

Motor over current (rate current) - Reduce pump motor speed

- Use fluid with lower viscosity
- Check if the pump impeller is blocked

Err 03

Pump off
Heating/refrigerating off

Motor over current (Max current) - Reduce pump motor speed

- Use fluid with lower viscosity
- Check if the pump impeller is blocked

Err 04

Pump off
Heating/refrigerating off

Motor hall signal missing

- Reduce pump motor speed
- Use fluid with lower viscosity
- Check if the pump impeller is blocked

Err 05

Pump off
Heating/refrigerating off

Too high liquid level

- Check the liquid level and buoyage

Err 06

Pump off
Heating/refrigerating off

Too low liquid level

- Check the liquid level and buoyage

Err 07

Pump off
Heating/refrigerating off

Too high voltage

- Check the mains power

Err 08

Pump off
Heating/refrigerating off

Too low voltage

- Check the mains power

Err 09

Pump off
Heating/refrigerating off

Device internal temperature is 
too high

- Check the ambient temperature and let the 

device cool down

Err 10

Pump off
Heating/refrigerating off

PC communication failure 

- Check communication cable
- Check the fan and clean the grids at the rear side

Err 11

Pump off
Heating/refrigerating off

Temperature difference 
between control sensor and 
safety sensor is too much

- Check safety temperature circuit and bath 

fluid

Err 12

Pump off
Heating/refrigerating off

Safety temperature alarm

- Check the bath temperature measurement

Err 13

Pump off
Heating/refrigerating off

Heater switched off by safety 
circuit

- Check safety temperature set value, fluid 

level

Err 14

Pump off
Heating/refrigerating off

Fan error

- Restart the device

Err 15

Pump off
Heating/refrigerating off

Safety shutdown of the 
compressor

- Restart the device

Err 16

Pump off
Heating/refrigerating off

Errors at the compressor 
(suction side)

- Restart the device

Err 17

Pump off
Heating/refrigerating off

Errors at the compressor 
(pressure side)

- Restart the device

Err 20

Pump off
Heating/refrigerating off

Fan cooling system error

- Restart the device

Err 21

Pump off
Heating/refrigerating off

Internal  error

- Restart the device

If the actions described fails to resolve the fault or another error code is displayed then take one of the following steps: 
- Contact the service department
- Send the device for repair, including a short description of the fault� 

Summary of Contents for CBC 5 basic

Page 1: ...20000013102 IKA CBC 5 basic CBC 5 basic_112016 Operating instructions EN...

Page 2: ...er 4 Handle 5 Venting grid 6 Mains switch 7 Power socket 8 Cable clip 9 Pump connection IN 10 Pump connection OUT 11 RS 232 interface 12 USB interface 13 External temperature sensor socket Device setu...

Page 3: ...panel and display 10 Setting the safety temperature 11 Commissioning 11 Menu settings 12 Menu structure 12 Basic guide to setting the menu options 13 Operating mode MODE 13 Maximum temperature HI T 13...

Page 4: ...10 2 051 EN 61326 1 EN 60529 and DIN 12876 1 Indicates an extremly hazardous situation which if not avoided will result in death serious injury Indicates a hazardous situation which if not avoided can...

Page 5: ...For example users can implement measures that inhibit fire or explo sions or comprehensive monitoring equipment Process pathogenic material only in closed vessels under a suitable fume hood Please co...

Page 6: ...at also supplies residential areas The safety of the user cannot be guaranteed if the device is operated with accessories that are not sup plied or recommended by the manufacturer if the device is ope...

Page 7: ...ources and do not expose to sun light Cooling machine pump motor and electronics produce intrinsic heat that is dissipated via the venting grids 5 Never cover these venting grids Note After setting up...

Page 8: ...valve is closed Ro tate clockwise to the stop position see Fig 7 Note Please note information in chapter Commissioning Connect the mains plug and turn on the device with mains switch 6 The low level...

Page 9: ...distilled or bi distilled water are unsuitable for operation due to corrosive properties of these media High purity water and distillates are suitable as a medium after adding 0 1 g soda Na2 CO3 sodi...

Page 10: ...speed value Start stop the pump function K Rotating pressing knob Set the temperature value Start stop the heating refrigerating function Navigation selecting the settings in the menu Set and confirm...

Page 11: ...and fan starts to run at once After pressing the mains switch 6 the start up of the com pressor is delayed about 3min In working status press the right knob J to stop the pump function The heating ref...

Page 12: ...play will change back and show pump speed When the heating refrigerating function is activated you can enter the menu option by pressing menu key G about 3 seconds Menu settings MODE Maximum temperatu...

Page 13: ...luid fluid type Note The maximum value is limited by the set safety temperature Minimum temperature LO T The minimum adjustable value 25 C Maximum speed HI R The maximum adjustable value 4600 rpm Mini...

Page 14: ...not displayed in the menu list The sampling time EuT 2 s is the interval over which the external control deviation and the respective control variable dependent on Kp 2 Ti 2 Td 2 Df t and FAST is dete...

Page 15: ...cs of the interface and the al location of signal status standard RS 232 applies in accor dance with DIN 66 259 Part 1 Transmission procedure asynchronous character transmis sion in start stop mode Ty...

Page 16: ...ODE_1 Set to External temperature control OUT_WD1 n Start the watchdog mode 1 and set the time for the watchdog to n 20 1500 seconds Echo of the Watchdog time During a WD1 event the heating and pump f...

Page 17: ...erial Water containing tenside isopropyl alcohol Cosmetics Water containing tenside isopropyl alcohol isopropyl alcohol Water containing tenside Foodstuff Water containing tenside Fuel Water containin...

Page 18: ...off Heating refrigerating off Device internal temperature is too high Check the ambient temperature and let the device cool down Err 10 Pump off Heating refrigerating off PC communication failure Chec...

Page 19: ...be nominal width 12 mm H PUR 8 PUR tube nominal width 8 mm H PUR 12 PUR tube nominal width 12 mm H FKM 8 FKM tube nominal width 8 mm H FKM 12 FKM tube nominal width 12 mm Tubing insulations ISO 8 Insu...

Page 20: ...play LED Heating capacity W 2500 230 VAC 1250 115 VAC Cooling capacity according to DIN 12876 at 4600 rpm 20 C 10 C 0 C 10 C 20 C W 350 400 W at 3200 rpm 320 370 W at 3200 rpm 270 320 W at 3200 rpm 19...

Page 21: ...IKA warranty conditions the warranty period is 24 months For claims under the warranty please contact your local dealer You may also send the machine direct to our factory enclosing the delivery invo...

Page 22: ...IKA Werke GmbH Co KG Janke Kunkel Str 10 D 79219 Staufen Tel 49 7633 831 0 Fax 49 7633 831 98 sales ika de www ika com...

Reviews: