Operator panel and display
Item Designation
Function
A
On/Off key:
Switch on/off the circulator�
B
LED, set:
The LED lights up at the same time as the set value is displayed�
C
LED, temperature sensor:
Indicates that an external temperature sensor is connected�
D
LED, heater:
Indicate the heating function is activated�
E
LED
,
refrigerating system:
Indicates that the refrigerating function is deactivated�
F
LED display:
Display the settings and actual values�
G
Menu key:
Press it once: menu option is display�
Press it a second time: back to the working screen�
H
LED
,
lock:
Indicates that the function of keys and knobs is deactivated�
I
Lock key:
Lock/unlock keys and knobs�
J
Rotating/pressing knob:
Set the pump speed value�
Start/stop the pump function�
K
Rotating/pressing knob:
Set the temperature value�
Start/stop the heating/refrigerating function�
Navigation, selecting the settings in the menu�
Set and confirm the menu options�
L
Adjustable safety circuit:
Adjust the safety temperature limit�
• Moving the device:
Empty all fluid in the bath before moving device from one
place to other place�
The device must be lifted by two persons with the upper
handles� For this purpose, take off the front cover (3)�
It can also be moved on flat surface by lifting and pushing
the front of the device� It is easy to move the device with
the help of the wheels�
The angle of inclination should never be more than 10 º
at any direction when move the device!
Note:
The device must not be moved during operation�
After moving the device, you must wait at least one hour
before restarting the unit�
Fig. 10
K
J
A
L
C
B DE
F
G
H I
max. 10 ˚
Fig. 9