Working with the sample rack
To work with the sample rack you must first check
that it is connected to the calorimeter.
Open the “Sample rack” screen from the main
menu.
After this you can use the „Use sample rack“
menu item to activate/ deactivate its use.
A schematic representation of the assignment
of the rack is displayed in the upper area of the
menu.
-> Green = free
-> Red = occupied
-> Grey = rack not connected/recognised
Working with the sample rack and balance
If a balance is connected to the
IKA
®
C 6000
global standards/isoperibol
calorimeter, the
weighted sample can be sent directly from the
balance to the
IKA
®
C 6000 global standards/
isoperibol
calorimeter.
There are two possibilities:
a) Press the “Print” button on the balance:
Pressing the “Print” button automatically opens
the preparation dialogue box and the
weighted sample value is entered automatically.
If External Energy 2 is not set to user-defined, the
first weighted sample value is entered into the
weighted value field for External Energy 2 and the
energy is calculated automatically.
If External Energy 2 is set to user-defined, the
weighted sample value is entered directly into the
test weighted value field and the second weigh-
ted sample value is entered into the External Ener-
gy 3 field (as long as it is not set to user-defined).
The energy values are calculated automatically
based on the weighted sample and
reference calorific value entered.
The sequence of figures transferred is as follows:
1. Weight External Energy 2 (if not user-defined)
2. Test weighted sample
3. Weight External Energy 3 (if not user-defined)