45
Empty Water:
The inner vessel can be emptied manually.
Stirrer:
Here the operation of the stirrer can be tested.
After the start the stirrer speed is displayed.
check cooling water:
After the start the current cooling water tempera-
ture at the entry into the calorimeter is displayed.
Cover up/down:
The cover can be opened and closed manually.
RFID:
After the start the decomposition vessel detec-
tion is activated and the RFID code of the decom-
position vessel can be read.
Click on „Stop“ to stop the function test.
Hardware details
In addition you can switch to the „Hardware de-
tails“ tab.
This tab shows you all the information about the
digital inputs and outputs, the temperature va-
lues on all the measurement channels, various
counters and statuses of accessories. You can-
not make any changes here. The information
displayed can be forwarded to Service Engineers
who have access to various settings.
Info
Here you can call up important information for
servicing purposes:
- Serial number of the calorimeter
- Software version
- PCB* number
- PCB* works number
- Number of ignitions
*PCB = printed circuit board