35
For management of the decomposition vessels
and their associated settings, click on the „Menu“
button and then the menu item „Function, de-
composition vessels and calibration“.
This list shows all the decomposition vessels as-
signed to the calorimeter. The mark on the left
in each line shows whether the decomposition
vessel is actually being used. For reasons of safety
decomposition vessels cannot be deleted.
Each decomposition vessel is uniquely identified
by its RFID tag. In addition the C values for the
current function, the date of last calibration and
the number of ignitions are displayed.
Click on the „Check calibration for selected de-
composition vessels and functions“ button. Three
additional tabs appear listing the calibration, a
check chart of the calibration and the calibration
statistics for the selected function and decompo-
sition vessel.
You can select the function of the
IKA
®
C 6000
global standards/isoperibol from a list.
Working Mode, Vessels and Calibrations
Alternatively you can use the actions provided by
the menu as described above.
Once you have completed the work on the mea-
surements for the selected day, switch back to
the „Library” tab.
Once you have completed all the library work,
click on the „Back“ button.
IKA
®
C 6000 global standards
IKA
®
C 6000 isoperibol