11
11
Unpacking
•Unpacking
- Please unpack the device carefully
- In the case of any damage a detailed report
must be set immediately (post, rail or forwar-
der)
• Delivery scope
RO
RT
- Magnetic stirrer
- Heating magnetic-
RO5/10 or 15
stirrer RT5/10 or 15
- power supply unit
- mains cable
- operating instruction
- operating instruction
Correct use
• Use
- For mixing and / or haeting liquids
• Range of use
(inside only)
- Laboratories
- Schools
- Pharmacies
- Universities
The device is suitable for use in all areas except:
- Residential areas
- Areas that are connected directly to a low-voltage supply net-
work that also supplies residential areas.
The safety of the user cannot be guaranteed if the appliance is
operated with accessories that are not supplied or recommended
by the manufacturer or if the appliance is opreated improperly
contrary to the manufacturer‘s specifications or if the appliance or
the printed circuit board are modified parties.
After the device is switched on a display self-test is performed.
The display cycles through the following information:
• All segments light up
• The software version and the selected mode are displayed
• The selected “Eco Mode” is displayed
(E when ECO mode is active)
• If the drive has not been started the display reads „OFF“. If
the drive has been started the selected speed is displayed.
Commissioning
RO/RT
Observe the ambient conditions (temperature, humidity, etc.)
listed under Technical Data.
The device is switched on and off via the switch (A) located on
the right-hand side of the device. On the front panel a small
arrow labelled Power above the display (B) indicates the location
of the switch.
Safe temperature limit
RT
The max. attainable heating plate temperature is restricted to
130°C by a permanently set safety temperature limit. Once this
limit has been attained, the device switches off the heating.
The safe temperature limit must all-
ways be set at least 25 °C lower then
the flash point of the media to be
processed!
Setting the safe temperature limit
RT
The safety temperature is set as follows:
Switch the device on using the On/Off switch
on the right side of the device.
As soon as the display shows SAFE, press and
hold down the Temp membrane key and use
the +/- membrane keys (
H
or
I
) to set the
desired safety temperature.
The set safety temperature is displayed.
Once the Temp membrane key is released, the safety temperature
is saved.
RO/RT
The stirring function is started and stopped by pressing the „Start/
Stop“ button
(C).
Pressing the +/- buttons
(E or D)
increases or reduces the speed.
Stirring function
WARNING