Reset to factory settings
Resets all settings to the state on delivery. Before the fac-
tory reset check the possible effects on your application!
• Press the "Mode/Enter" button until rES is displayed
• Press the "Set" button and keep it pressed until
"
- - - -
" is displayed,
• Press the "Mode/Enter" button briefly.
The display then returns to
rES
.
Setting range
for the
probe length
(applies to all LR70 type units)
The probe can be replaced. Probes are available as accessories in different
lengths.
Setting range
for
OFS
(applies to all LR70 type units)
Setting range
for
SPx
,
rPx
*L = probe length [cm] on delivery;
∆
L = increments
The values apply to the probe length on delivery and OFS = 0.
•
rPx is always smaller than SPx. Reducing the value for SPx to a value
≤
rPx
also shifts the position of rPx.
•
If the difference between rPx and SPx is small (approx. 3 x increment),
increasing rPx also increases SPx.
•
If the difference between rPx and SPx is great, rPx remains at the set value
38
Parameter
Step
Programming
Setting range
in steps of
cm
15...70
0.5
inch
6.0...27.6
0.2
Setting range
in steps of
cm
0...100
0.5
inch
0...39.4
0.2
SPx
∆
L*
LR7022 ( L = 24)*
cm
inch
1.5...20.0
0.6...7.8
0.5
0.2
rPx
1.0...19.5
0.4...7.6
LR7023 (L = 45)*
cm
inch
1.5...41.0 0.6...16.2
0.5
0.2
1.0...40.5 0.4...16.0
LR7024 ( L = 70)*
cm
inch
1.5...66.0 0.6...26.0
0.5
0.2
1.0...65.5 0.4...25.8