Image Diagnostics Inc
Aspect100UC Plus
Mobile Urology Table
Operator Manual
Made in the USA
Page 1: ...Image Diagnostics Inc Aspect100UC Plus Mobile Urology Table Operator Manual Made in the USA...
Page 2: ...1 SAFETY INSTRUCTIONS 13 ACCESSORIES 15 EMERGENCY STOP SWITCH 26 ELECTRIC SYSTEM 27 CONTROLLER CONNECTIONS 28 GENERAL CLEANING 29 MAINTENANCE 32 USER REPLACEMENT ITEMS 33 TROUBLE SHOOTING 33 DISPOSAL...
Page 3: ...x 86cm Longitudinal travel 10in 25 4cm Trendelenburg 15 Lateral travel 3 5in 88mm Height range 30in to 40in 76cm to 101cm Patient capacity 525lbs 238kg including additional accessories without tableto...
Page 4: ...This equipment must be installed in an area provided with the proper electrical power The owner of this equipment is responsible for verifying continued compliance with all applicable regulations and...
Page 5: ...s calibration instructions or other information to assist the user s appropriately qualified technical personnel to repair those parts of equipment which are designated by the manufacturer as repairab...
Page 6: ...age to equipment CD ROMs containing file copies of all relevant drawings BOMs and documentation are included with this manual one for general operators and one for the BIOMED department Type B equipme...
Page 7: ...tem This symbol when used in this manual indicates the potential of exposure to harmful x rays Be sure to read and comply with all warnings European Authorized Representative Warning Information or in...
Page 8: ...0 10 Vac 50Hz Duty Cycle 10 6 min hr Current Rating Less than 10 Amps Internally Battery Powered Environmental Operating Temperature Range 10 C to 40 C Operating Humidity Range 30 to 75 relative humid...
Page 9: ...pment must be installed and used according to the manufacturer s instructions in order to avoid radio frequency interference If this equipment generates or receives interference Verify that the equipm...
Page 10: ...oduct and before placing table into service table condition and functionality should be reviewed by a Biomed or equivalent in house support service resource Install Mobile tables are equipped with fou...
Page 11: ...power cord from the mains Preparation for Patient Use New Installation This equipment will need to be properly cleaned before patient use as it will inevitably come in contact with contaminants durin...
Page 12: ...Indicates an imminently hazardous situation which if not avoided will result in death or serious injury WARNING Indicates a potentially hazardous situation which if not avoided could result in death o...
Page 13: ...constitutes potential expose to harmful x rays for both the patient and operator Be sure to use proper radiation shielding CAUTION Secure power cord when in use so that it does not get entangled with...
Page 14: ...al cross travel position when loading or unloading patient Always safely position and secure patient on table WARNING Do not exceed patient weight of 400lbs 181kg when the tabletop extension IS attach...
Page 15: ...nd a right boot CAUTION Do not sit at the head of the table or on the table extension CAUTION Do not place or store any containers or large items underneath the tabletop As the tabletop is descending...
Page 16: ...the extension onto the tabletop accessory rails with the extension held at a slight angle as shown Bring the end of the extension level as the extension fully engages Try to pull the extension straigh...
Page 17: ...www imagediagnostics com 16 EXTENSION REMOVAL Tilt the far end of the extension up and lift off Tabletop Extension Part Number A100 1400 Patient Restraint Straps Hook loop fasteners Patient Restraint...
Page 18: ...and left supports are not interchangeable INSTALLATION 1 Line up the large and small pin to the holes in the mounting block located at the perineal end of the table 2 Slide pins into appropriate hole...
Page 19: ...ND INSTALL STIRRUP TABLETOP EXTENSION CAN SLIDE OFF ON WITH STIRRUPS IN PLACE The IDI Stirrup Extension Kit allows the stirrups to be attached to the table in an advantageous position that allows the...
Page 20: ...t way to keep instruments conveniently close to the doctor These have pins that drop into the holes that the elbow supports would use The trays can t be used at the same time as the elbow supports The...
Page 21: ...tion D Hook loop patient straps Patient Armboard with Pad Restraint Straps Part Number A310 056 see note below Replacement Armboard Pad Part Number X100 1755 see note below Replacement Armboard Strap...
Page 22: ...henever the table is used for procedures involving the use of a mobile C arm or any other equipment that produces ionizing radiation all radiation safety standards and precautions should be applied in...
Page 23: ...bag support A small pins down through opening in drain bag edge B Insert rest of support until bag is centered 3 Align small pins of support with support blocks C Push in until pins are firmly locked...
Page 24: ...Pull plastic bag edges off the support pins 2 Pull out drain bag support One Time Use Only Not made with natural rubber latex DISPOSAL Per local procedures typically normal waste or incinerate Drain B...
Page 25: ...hed The tabletop will pause at the horizontally level position during Trendelenburg motion Release and activate control again to continue past level As the position of the patient s head may be at eit...
Page 26: ...ition of the patient s head may be at either end of the table the action may be opposite to what is indicated by the label on the control Use foot control attached to table or place on floor Foot cont...
Page 27: ...ith natural rubber latex COVER DISPOSAL Per local procedures Typically normal waste or incinerate Foot Control Disposable Cover Foot Control Part Number C000 0492 Box of 50 Part Number A100 1540 Foot...
Page 28: ...l malfunctions the emergency stop control prohibits motion by removing power from all system components ACTIVATE Engage the emergency stop mode by pushing the button down RESET Restore the electrical...
Page 29: ...y fuse inside B Safe DC Voltages C Level Position Switch D Trendelenburg Tilt Actuator E Elevating Lifting Actuator F Slide Longitudinal Actuator G Lateral Cross Actuator in line fuse in cable H AC Po...
Page 30: ...automatically switch to battery back up mode Local procedures may be in place for proper disposal or recycling For more information about batteries such as battery life production codes and disposal g...
Page 31: ...f 5 25 sodium hypochlorite diluted between 1 10 and 1 100 with water Alcohol generic Envirocide Disinfectant and Cleaner CLEANING STEPS 1 Move the tabletop into the horizontal position 2 Lower the tab...
Page 32: ...an adequately ventilated area 2 Apply to one side at a time and allow to stand a few minutes 3 Scrub with a soft bristle scrub brush Do not let the solution dry before rinsing 4 Rinse with water and a...
Page 33: ...ater 2 Rinse with water and dry or to clean and or disinfect 1 Use a diluted bleach solution of 10 parts water to 1 part bleach 2 Wipe dry with a clean cloth APPROVED AND TESTED DISINFECTANT CLEANER L...
Page 34: ...CKS Daily inspect all external cables controls and the tabletop for wear and damage Replace damaged cables promptly Check daily if table is routinely used without being attached to mains Weekly check...
Page 35: ...s 94 UL94 The vinyl coating is self extinguishing when free of solvents LATEX Not made with natural rubber latex Tabletop Pad 2in x 28in x 46 1 2in 5cm x 71cm x 118cm Part Number X100 1818 Extension P...
Page 36: ...ts components typically found in IDI products but varies with model and options COMPONENT ITEM RECYCLING GROUP Actuator Spindle and Motor Housing Cable Metal Steel and Copper Plastic Copper Control Bo...
Page 37: ...L100 3065 REV C www imagediagnostics com 35 WARRANTY Warranty details for IDI Products can be obtained directly from Image Diagnostics Inc Image Diagnostics Inc 310 Authority Drive Fitchburg MA 01420...