Technical specifications
Sample volume: 165 μL
Database capacity: up to 1,000 samples
Interfaces: Ethernet
Dimensions
Width: 12.6 in. (320 mm)
Depth: 12.6 in. (320 mm)
Height: 13.2 in. (336 mm)
Weight: 32 lb (15 kg)
Power requirements
Input: 100–240 V AC
Line frequency: 50–60 Hz
Operating conditions
Indoor use only
Ambient room temperature: 15°C–40°C (59°F–104°F)
Relative humidity: 20%–80% at 30°C (86°F), noncondensing
Altitude: Up to 2,000 meters above sea level
Safety precautions
IMPORTANT:
If the equipment is used in a manner other than specified, the protection provided by
the equipment may be impaired.
The analyzer does not contain any user-serviceable components. DO NOT disassemble.
Line voltage for the SediVue Dx AC power cable is 100–240 V AC, 50–60 Hz. Be sure to plug all
equipment into properly grounded electrical outlets.
Use only the AC power cable supplied.
Disconnect the AC power cable from the wall outlet if the:
• AC power cable becomes frayed or otherwise damaged.
• AC power adapter is exposed to water or other liquids.
Electrical safety
Removing the covers of electronic equipment can cause electric shock, as there are high-voltage
parts inside.
Do not tamper with any of the electronics inside the analyzer.
Do not remove any analyzer cover other than those specified in this document.
To prevent fire or shock hazard, do not expose the analyzer to rain or moisture of any kind.
To prevent electric shock, never use an extension cord. Ensure all plugs and sockets are fully
inserted so that no parts of the plug prongs or blades are visible.