26
6b. Using a new alcohol
prep pad, clean the
optical window in a
circular motion (3–4
times) and then dry
it in a circular motion
with a clean optical
tissue.
7. Remove the centrifuge shield, waste bin and drip tray. Then clean
them with warm water and dry them with an optical tissue.
8. Using a new alcohol
prep pad, clean the
move arm and then
dry it with an optical
tissue.
Move arm
9. Return the waste bin,
drip tray, centrifuge
shield, and cartridge
holder to their
locations in the
analyzer.
10. Close the analyzer
door. On the IDEXX
VetLab Station Home
screen, tap the
SediVue Dx
icon
and then tap
Turn
on SediVue Dx
.
Maintenance
Optical
window