2 Operational Check
3 Window and Menu
4 Appendix
1 Overview
4-2
Troubleshooting
4. 1. 3
Cannot connect to the WB1F Support Tool because the WB1F RS-232 settings were forgotten
On the Connection Settings screen, select the communication port to which the WB1F is connected, and then
click Automatic connection.
The support tool automatically searches for the communication settings that the WB1F is set to.
This may take some time. (The maximum time: About 1 min 30 sec.)
Check the communication settings from maintenance mode.
1
Switch to maintenance mode.
(Refer to the WB1F user’s manual.)
2
Start the WB1F Support Tool and click Connection with the default settings.
• Default settings
Communication speed
9600 bps
Flow control
None
Data length
8
Prefix
5E,00,00,00
Parity
EVEN
Suffix
0D,0A,00,00
Stop bits
1
Check digit
None
3
Double-click Console to open the Console screen.
4
Enter “comgetm” and click Send.
* Select only the Prefix ON and Suffix ON check boxes.
5
Check the communication settings that are set on the WB1F from the response results (red).