Ensuring adequate ventilation

 

10.0  There must be good ventilation surrounding the 

ICS PACIFIC 

G135L Medical 

fridge. Insufficient ventilation will cause the fridge to run outside the required 
operating temperature range and can cause permanent and expensive damage to 
both the compressor and vaccines. 

For further information please contact the 

ICS PACIFIC

 service department. 

 

Hi/Lo Temperature Light

 

11.0  When this indicator is illuminated the display is currently showing the High or Low 

temperature reading. The reading is not indicative of the current temperature. 

 

Refrigerating Light

 

12.0  This indicator may flash as part of its normal operation (It has nothing to do with the 

temperature inside the fridge). 

 

Defrost Light

 

13.0  This indicator shows the fridge is in defrost. The fridge will defrost every 6 hours. 

This indicator will flash after defrost as part of its normal operation. 

 

The UP/Light Button

 

14.0  The UP/Light button switches the internal light on and off.  
 

The DOWN/Mute Button

 

15.0  The DOWN/Mute button is used to deactivate the alarm in the event it is activated.  
 

The HI/LO Button

 

16.0  The Hi/Lo Button is used to check the maximum and minimum temperatures of the 

fridge. It is also used to clear and reset these temperatures. 

 
 

The Alarm Function 

17.0  The alarm will only operate if the cabinet’s temperature reaches above +8ºC or below 

+2ºC for more than 30 minutes.  

 

If the cabinet resumes normal operation, the alarm will automatically deactivate.  
 

Note:

 The Hi/Lo function will record any high or low temperature breach. 

 

 

 

 

 

 

 

 

 

Toll Free 1800 630 360

 and quote the serial number of the fridge. 

Please note: Please be aware that if a technician is sent and no 

warranty fault is found there will be a service charge. 

Summary of Contents for G135L

Page 1: ...anual 31 Healey Road Dandenong South Victoria 3175 Australia P O Box 4140 Dandenong South BC Victoria 3164 Australia E mail info icspacific com au Web www icspacific com au Phone 61 03 9706 5115 Facsi...

Page 2: ...cabinet TABLE OF CONTENTS INTRODUCTION 2 SAFETY INSTRUCTIONS 3 SAFETY INSTRUCTIONS 3 INSTALLATION STOCKING INSTRUCTIONS 4 SETTING UP THE CABINET 4 INSTALLING THE HANDLE 4 INSTALLING THE BOTTOM SHELF...

Page 3: ...e cabinet off at the power point first and remove the 3 pin plug carefully from the power point Always disconnect the 3 pin plug of the cabinet from the power point before cleaning to avoid danger of...

Page 4: ...wn dedicated power point Do not use an extension cord double adaptor or power board INSTALLING THE HANDLE Cut the cable tie holding the handle in place Fit the handle to the left hand side of the door...

Page 5: ...wall Before cleaning the cabinet remove all products from within the cabinet When cleaning the cabinet interior exterior it is recommended that you use a solvent of warm water and mild detergent or a...

Page 6: ...LO Button and hold it in for 6 to 7 seconds The reset will be signalled by one long beep 4 Every time the high and low temperatures are checked it is important to clear and reset the displayed values...

Page 7: ...icator shows the fridge is in defrost The fridge will defrost every 6 hours This indicator will flash after defrost as part of its normal operation The UP Light Button 14 0 The UP Light button switche...

Page 8: ...p within the evaporator Ensure that the drain tube has not dislodged from the drip tray at the rear of the cabinet If there is condensation on the cabinet The ambient room temperature must not be abov...

Reviews: