ICS Schneider Messtechnik PPS 1210 Operating	 Instruction Download Page 9

T

echnical data

 

 

 

 

Pressure range (FS) 

 

0…700 bar   (others on request)

 

Overpressure 

 

 

840 bar (FS+20%)

 

Resolution of display 

 

1mbar 

 

Accuracy, error band (1) (10…40 °C) 

< 0,05 %FS

 

Accuracy, error band (1) (0…50 °C) 

< 0,1 %FS

 

Leak rate* 

 

 

700 bar: -2 bar @ 10 min.

 

Number of digits on display   

5 digits

 

Measurement interval 

 

0,5 seconds

 

Interface 

 

 

USB (RS485 out of function)

 

 

Compensated temperature range 

0…50 °C

 

Operating temperature 

 

0…50 °C

 

Storage temperature 

 

-10…60 °C

 

Air humidity 

 

 

5…95% relative humidity

 

 
Power supply  
heatable pressure port and sensor 

115V AC or 230V AC 

Display 

 

 

Button-cell battery, type CR2430

 

Battery lifetime   

> 2000 h in continuous operation

 

Hydraulic oil 

 

 

HLP 22 BP or olive oil 

 
Pressure port 

 

 

1/2"20UNF or M18x1,5 

Media compatibility 

 

Pressure transducer with stainless steel diaphragm 

 

Dimensions (L x W x H) 

 

470 x 170 x 280 mm 

Weight 

 

 

approx. 7,2kg

 

Degree of protection 

 

IP 54

 

 
 

Selectable pressure units 

 

bar, mbar, hPa, kPa, MPa, PSI, kp/cm2, cmH2O, mH2O, inH2O, ftH2O,

 

 

 

 

 

mmHg, inHg 

 
 

Delivery in a stable aluminium transport case!

 

 

 
 
 
 

 

(1) 

including accuracy, temperature coefficients, zero point and range tolerance

 

 

* Physical effects caused by a pressure change lead at first to a clear difference in pressure.

 

Advice: To minimise the influence of these physical effects increase steadily the last 5% of the target pressure and regulate 

towards the target pressure for the first minutes.

 

 

The stated leakage rate is at a thermal balanced condition (when temperature of pressure media and of the environment is equable).

 

 
 
 
 
 
 
 

 

Summary of Contents for PPS 1210

Page 1: ...instruction high pressure calibrator PPS 1210 L ICS Schneider Messtechnik GmbH Briesestra e 59 D 16562 Hohen Neuendorf OT Bergfelde Tel 03303 504066 Fax 03303 504068 info ics schneider de www ics sch...

Page 2: ...1 Description of the device 4 2 General safety warnings 4 3 Operating the calibrator 5 4 Description of the functions 5 5 Menu navigation for calibrators 6 6 Commissioning 6 7 Maintenance disposal 7 8...

Page 3: ...ealing plug 8 Manual booster pump 9 Pressure connection for test object only PPS1110 10 Drain valve 11 Screw compressor Additional components of High Pressure Calibrator PPS 1210 12 Heatable pressure...

Page 4: ...IMPORTANT Nothing must adhere to the surface of the test object no oil grease water etc Impurities could pass trough the adapter to reach the high pressure calibrator and damage it Overpressure If th...

Page 5: ...each the desired pressure please consult the section on Maintenance to find out how to vent the system Zeroing the device Open the drain valve 10 to release any pressure that may have built up If the...

Page 6: ...6...

Page 7: ...ssure change Set leak measurement time The leak measurement time is preset to 10 minutes in the factory and it can only be changed with the Mano Con fig software Software for calibrators MANO Continuo...

Page 8: ...15 L 17 L 19 L 21 and L 23 are completed The candidate s 100 bar range is entered in the cell L 11 printing area The cells L 11 L 15 L 23 and P 31 are obligatory fields The result of the measurement...

Page 9: ...LP 22 BP or olive oil Pressure port 1 2 20UNF or M18x1 5 Media compatibility Pressure transducer with stainless steel diaphragm Dimensions L x W x H 470 x 170 x 280 mm Weight approx 7 2kg Degree of pr...

Page 10: ...Konformit tserkl rung Declaration of Conformity D claration de Conformit 10...

Page 11: ...al use thanks to the many important customisation options it offers These include useful additional functions such as sensor break detection with automatic power cutoff alarm output maximum or minimum...

Page 12: ...ity in Resolution 0 1 C Display resolution is customisable non aggressive environment Closed circuit current 3W Control Display elements Status message display 1 4 4 RTD The display then changes to op...

Page 13: ...OFF regulation is that the temperature fluctuates in the range around the set value due to the hysteresis that was specified and the sluggishness of individual process components heater follow up time...

Page 14: ...0 changes in menu A and autotuning are permitted LOC 0 Changes in Menu A are not permitted exception parameters LOC 508 Call up and changes to menu B are permitted After leaving Menu B Loc jumps back...

Page 15: ...temperature value is added to both ascending and descending temperature slopes This therefore doubles the total hysteresis switching gap Example of a heating control system on output relay terminal 1...

Page 16: ...nformation provided in menu table B under parameter point ALT1 If when the regulator is switched on an undesired alarm is emitted when the set value at the time of switching on corresponds to the alar...

Reviews: