Notes on the operating instructions

 

 

-  The  operating  instructions  are  in- 

tended  for  specialist  workers  and 
trained personnel.

 

-  Before each stage of work, read the 

relevant notes  and  warnings care- 
fully, and keep to the sequence as 
stated.

 

-  Pay particular attention to the sec- 

tion on ”General safety warnings”.

 

 

If you have any problems or questions, 
please  contact  your  supplier  or 
consult  the  company  MARIANNE 
MAYER, ELECTRONISCHE SCHAL-
TUNGEN  directly. 
 
1.  Description of the device

 

 

General description

 

The high  pressure  calibrator  enables 
pressure to be generated by means of  
the  integrated  pressure  pump,  up  to 
700 bar relative.

 

The  measurement  technology  incor- 
porated  into  this  device  allows  ac- 
curate  measurement  and  documen- 
tation  of  the  characteristic  of  a  test  
object  that  is  connected  to  it.  The 
measured  pressure  progression  can 
be  displayed,  evaluated  and  saved 
with  a  computer  monitoring  program 
(CCS30).

 

The  calibrator  is  operated  with  the 
two  function  buttons  SELECT  and 
ENTER,    located      directly      below 
the  display.  The  calibrator  itself  is 
powered by a 3,0 V battery.

 

 
Copyright

 

For  copyright  reaons  this  operating 
manual  may  only  be  used  for  inhouse 
purposes. 
Any  reproduction,  even  in  part  and  for 
inhouse  purposes,  requires  the  assent 
by 

Messrs 

 

Also,  any  distribution  to  third  parties  is 
prohibited for competitive reasons. 

Commissioning

 

A  pressure-resistant  connection  for 
the test object is required in order to

 

use the high pressure calibrator. The 
pressure connection for the test object 
is  already  screwed  to  the  pressure 
distributor of the high pressure calibra- 
tor so that it is pressure resistant when 
it leaves the factory, and it must not be 
dismantled.

 

Recommended  torque  for  the  test 
object pressure connection: 30 Nm

 

 

 

IMPORTANT!

 

Nothing must adhere to the surface of 
the test object (no oil, grease, water,

 

etc). Impurities could pass trough the 
adapter  to  reach  the  high  pressure 
calibrator and damage it. 

 

Overpressure

 

If the pressure exceeds the measuring 
range by more than 20%, the measur- 
ing cell or the mechanism of the high 
pressure calibrator may be destroyed 

 
 
Recalibration

 

The  recalibration  cycle  depends  on 
the conditions of use. Recommended 
recalibration cycle: 1 year.

 

 

Intended  use

 

The  high  pressure  calibrator  may 
only  be  used  to  generate  pressure 
with the type HLP 22 BP hydraulic oil 
that is supplied with the product. Use 
of the calibrator with other media

 

will 

damage it. The operational safety

 

of  the  device  supplied  is  guaranteed 
only if it is used as intended. The limit 
values as stated (see page 9: ”Tech- 
nical data”) must never be exceeded. 

 

Before  installing  the  high  pressure 
calibrator, check that it is suitable for  
your applications 

 

2.  General safety warnings

 

 

The  current  national  regulations  on 

accident 

 

prevention 

 

and  

workplacesafety  must  be  followed 
whenever  work 

is 

carried 

out. 

Internal  regulations  issued  by  the  
operator    must  be  followed,  even  if 
they  are  not  mentioned  in  these 
instructions.

 

 

Never  use  the  high  pressure  calibra- 

tor  together  with  an  external  pressure 
source.

 

 

Do not remove  any connected  compo- 

nents (e.g. test objects) when the high 
pressure  calibrator  is  under  pressure. 
Open the screwed sealing  plug before 
removing parts.

 

 

Do  use  Teflon    tape  to  seal    the  

pressure  connection careful.  Residues 
of  Teflon  tape  could  penetrate  the 
high-  pressure  calibrator  and  damage 
it.

 

 

Only  use  the  adapters  and  seals  that 

are available as accessories.

 

 

Do not store the calibrator under pres- 

sure: only store the high pressure cali- 
brator with the drain valve open.

 

 

Avoid  the  action  of  force  of  any  kind 

on the high pressure calibrator and its 
operating controls.

 

 

Do not  use  high  pressure calibrators if 

they are damaged or faulty.

 

 

4

 

Summary of Contents for PPS 1210

Page 1: ...instruction high pressure calibrator PPS 1210 L ICS Schneider Messtechnik GmbH Briesestra e 59 D 16562 Hohen Neuendorf OT Bergfelde Tel 03303 504066 Fax 03303 504068 info ics schneider de www ics sch...

Page 2: ...1 Description of the device 4 2 General safety warnings 4 3 Operating the calibrator 5 4 Description of the functions 5 5 Menu navigation for calibrators 6 6 Commissioning 6 7 Maintenance disposal 7 8...

Page 3: ...ealing plug 8 Manual booster pump 9 Pressure connection for test object only PPS1110 10 Drain valve 11 Screw compressor Additional components of High Pressure Calibrator PPS 1210 12 Heatable pressure...

Page 4: ...IMPORTANT Nothing must adhere to the surface of the test object no oil grease water etc Impurities could pass trough the adapter to reach the high pressure calibrator and damage it Overpressure If th...

Page 5: ...each the desired pressure please consult the section on Maintenance to find out how to vent the system Zeroing the device Open the drain valve 10 to release any pressure that may have built up If the...

Page 6: ...6...

Page 7: ...ssure change Set leak measurement time The leak measurement time is preset to 10 minutes in the factory and it can only be changed with the Mano Con fig software Software for calibrators MANO Continuo...

Page 8: ...15 L 17 L 19 L 21 and L 23 are completed The candidate s 100 bar range is entered in the cell L 11 printing area The cells L 11 L 15 L 23 and P 31 are obligatory fields The result of the measurement...

Page 9: ...LP 22 BP or olive oil Pressure port 1 2 20UNF or M18x1 5 Media compatibility Pressure transducer with stainless steel diaphragm Dimensions L x W x H 470 x 170 x 280 mm Weight approx 7 2kg Degree of pr...

Page 10: ...Konformit tserkl rung Declaration of Conformity D claration de Conformit 10...

Page 11: ...al use thanks to the many important customisation options it offers These include useful additional functions such as sensor break detection with automatic power cutoff alarm output maximum or minimum...

Page 12: ...ity in Resolution 0 1 C Display resolution is customisable non aggressive environment Closed circuit current 3W Control Display elements Status message display 1 4 4 RTD The display then changes to op...

Page 13: ...OFF regulation is that the temperature fluctuates in the range around the set value due to the hysteresis that was specified and the sluggishness of individual process components heater follow up time...

Page 14: ...0 changes in menu A and autotuning are permitted LOC 0 Changes in Menu A are not permitted exception parameters LOC 508 Call up and changes to menu B are permitted After leaving Menu B Loc jumps back...

Page 15: ...temperature value is added to both ascending and descending temperature slopes This therefore doubles the total hysteresis switching gap Example of a heating control system on output relay terminal 1...

Page 16: ...nformation provided in menu table B under parameter point ALT1 If when the regulator is switched on an undesired alarm is emitted when the set value at the time of switching on corresponds to the alar...

Reviews: