8. Evaluation with the PC 
 
Installation 
 
Before the first opening of the Excel sheet, copy the file "s30c.dll" in die folder C:\WINDOWS/SYSTEM (not SYSTEM32). 
You require administrator rights in order to be able to do this. If necessary, inform your IT administrator. 
 
Measuring and completing the Excel sheets 
 
Connect the USB output of the calibrator with the PC's USB port. 
Under Connections in the Device Manager, look for the number of the USB port of the PC, to which the calibrator is 

connected.  

Enter this value in the cell P-31 at the description "Com-Port-N" in the Excel sheet. 
In the next step, the cells E-16, L-9, L-11, L-15, L-17, L-19, L-21 and L-23 are completed. The candidate's 100% bar range 

is entered in the cell L-11 (printing area). 

 
The cells L-11, L-15, L-23 and P-31 are obligatory fields.  
 
The result of the measurement is evident from the cells G-26, G-27 and G-28. This information relates to the data in cell L-

15 (accuracy). 

 
Start the measuring by clicking the button "Measurement 1" in the depressurized state. 
The data of the reference sensor in the calibration and the candidate's are read. At the measurement 1, the calibrator's 

data with the serial number and the date, with the time of the measurement, is also automatically read. 

Increase the pressure of the calibrator to about 50% of the final value. Wait until the value stabilizes (evident on the 

calibrator's display). Click the button "Measurement 2". 

Increase the pressure in the calibrator to 100% of the candidate's, but only to a maximum of 700 bars.  
Caution:  The calibrator can be damaged or destroyed with a pressure of more than 700 bars (+ 10% max). 
Save the test protocol under a different name. 
 
The read data is deleted by clicking the "Delete All" button. 

 

 
 

 

Image:  Excel sheet for the evalution of the measurement 

8

Summary of Contents for PPS 1210

Page 1: ...instruction high pressure calibrator PPS 1210 L ICS Schneider Messtechnik GmbH Briesestra e 59 D 16562 Hohen Neuendorf OT Bergfelde Tel 03303 504066 Fax 03303 504068 info ics schneider de www ics sch...

Page 2: ...1 Description of the device 4 2 General safety warnings 4 3 Operating the calibrator 5 4 Description of the functions 5 5 Menu navigation for calibrators 6 6 Commissioning 6 7 Maintenance disposal 7 8...

Page 3: ...ealing plug 8 Manual booster pump 9 Pressure connection for test object only PPS1110 10 Drain valve 11 Screw compressor Additional components of High Pressure Calibrator PPS 1210 12 Heatable pressure...

Page 4: ...IMPORTANT Nothing must adhere to the surface of the test object no oil grease water etc Impurities could pass trough the adapter to reach the high pressure calibrator and damage it Overpressure If th...

Page 5: ...each the desired pressure please consult the section on Maintenance to find out how to vent the system Zeroing the device Open the drain valve 10 to release any pressure that may have built up If the...

Page 6: ...6...

Page 7: ...ssure change Set leak measurement time The leak measurement time is preset to 10 minutes in the factory and it can only be changed with the Mano Con fig software Software for calibrators MANO Continuo...

Page 8: ...15 L 17 L 19 L 21 and L 23 are completed The candidate s 100 bar range is entered in the cell L 11 printing area The cells L 11 L 15 L 23 and P 31 are obligatory fields The result of the measurement...

Page 9: ...LP 22 BP or olive oil Pressure port 1 2 20UNF or M18x1 5 Media compatibility Pressure transducer with stainless steel diaphragm Dimensions L x W x H 470 x 170 x 280 mm Weight approx 7 2kg Degree of pr...

Page 10: ...Konformit tserkl rung Declaration of Conformity D claration de Conformit 10...

Page 11: ...al use thanks to the many important customisation options it offers These include useful additional functions such as sensor break detection with automatic power cutoff alarm output maximum or minimum...

Page 12: ...ity in Resolution 0 1 C Display resolution is customisable non aggressive environment Closed circuit current 3W Control Display elements Status message display 1 4 4 RTD The display then changes to op...

Page 13: ...OFF regulation is that the temperature fluctuates in the range around the set value due to the hysteresis that was specified and the sluggishness of individual process components heater follow up time...

Page 14: ...0 changes in menu A and autotuning are permitted LOC 0 Changes in Menu A are not permitted exception parameters LOC 508 Call up and changes to menu B are permitted After leaving Menu B Loc jumps back...

Page 15: ...temperature value is added to both ascending and descending temperature slopes This therefore doubles the total hysteresis switching gap Example of a heating control system on output relay terminal 1...

Page 16: ...nformation provided in menu table B under parameter point ALT1 If when the regulator is switched on an undesired alarm is emitted when the set value at the time of switching on corresponds to the alar...

Reviews: