ICS Schneider Messtechnik Anton Paar DMA 501 Instruction Manual And Safety Information Download Page 16

7 Performing a Measurement

16

E15IB001EN-N

7.2  Filling Sample

You can fill the sample with a syringe or with an 
optional peristaltic pump. See the manual of the 
peristaltic pump for instructions on its installation 
and use.

To achieve highly accurate measuring results, fill 
the sample into the measuring cell steadily and 
without bubbles.

TIP: 

Sample containing dissolved CO

2

 will cause 

bubbles in the measuring cell, which render the 
measurement results invalid. Degas your sample 
carefully as described in section 7.5.

Sample amount

If the measuring cell is clean and dry, you need 
approx. 2 mL of sample.

If you measure without cleaning and drying between 
samples, you will need a higher amount of sample, 
because you have to flush residues of the previous 
sample out of the measuring cell to avoid cross-
contamination.

Bubble detection – FillingCheck™ and 
U-View™ (live camera image)

FillingCheck™

 (see section 6.3.6) helps you detect 

bubbles during filling. Activate the function in the 
measurement settings.

Additionally check the filling visually by means of the 
built-in live camera (

U-View™

):

• Place the camera widget on the dashboard 

assigned to your measurement.

• Watch the live camera image on the dashboard 

during filling:
- Tap on the image to open the camera screen.
- Tap 

 to zoom in, tap 

 to zoom back out.

- You can drag the zoomed image left or right 

to examine all parts of the measuring cell.

- Tap 

 to close the camera screen.

The camera image will be saved with the measure-
ment results.

 

WARNING

Risk of injuries and fire by liquids leaking

Filling samples and cleaning liquids that the 
wetted parts are not resistant to will corrode the 
wetted parts. Sample leaking from corroded parts 
may cause serious injuries.
Before you fill any sample or cleaning liquid, in 
particular hazardous or flammable chemicals, 
into the instrument:
• Strictly follow all safety instructions concerning 

the use of chemicals and the use of flammable 
chemicals, see section 1.

• Make sure that all wetted parts are resistant to 

the filled-in liquid, see appendix A.3.
Consider also the wetted parts of installed 
options.

• Make sure that you have suitable cleaning 

liquids at hand for cleaning the measuring cell, 
see section 9.1.

NOTICE

Corrosive samples require special care

Samples with a moderate tendency to corrode 
borosilicate glass, such as strong alkali solutions 
(e.g. caustic soda), can be measured with the 
instrument. However:
• Remove corrosive samples immediately after 

measurement and rinse the measuring cell 
thoroughly.

• Check the validity of the adjustment more 

frequently than generally recommended.

• The measuring temperature for strong alkali 

solutions should not be higher than 20 °C 
(68 °F). Higher temperatures dramatically 
increase the speed of corrosion.

NOTICE

Proceeding in case of instrument leaks

If liquid leaks from the draining hole by the 
sample inlet:
• Perform a leak tightness test as described in 

section 3.2.2.

• If the leak tightness test fails, contact your local 

Anton Paar representative.

NOTICE

Some organic solvents such as toluene, xylene 
or acetone may etch the instrument housing. 
Usually this occurs due to spilled sample or 
cleaning liquids.
• Remove any spilled liquids immediately as 

described in section 10.2.

• Use the protective cover in the case that spilling 

occurs.

• Detach the syringe holder from the instrument 

housing and position it at the side of the 
instrument to prevent etching on the housing 
where the syringe holder is located.

Summary of Contents for Anton Paar DMA 501

Page 1: ...Instruction Manual and Safety Information DMA 501 DMA 1001 Density Meter instrument software version from 4 1 0 Original Instructions...

Page 2: ...and its contents may be changed or amended by Anton Paar at any time without prior notice All rights reserved including translation This document or any part of it may not be reproduced changed copie...

Page 3: ...13 6 Measurement Settings Products 14 6 1 Products 14 6 2 Measurement Settings 14 6 2 1 Dashboards 14 6 2 2 Reports 14 6 3 Measurement Parameters 14 6 3 1 Set Temperature 14 6 3 2 Measurement Perform...

Page 4: ...ed Anton Paar Service Engineer 25 10 2 Repair Performed by an Authorized Anton Paar Representative 25 Appendix A Technical Data 26 A 1 Specifications 26 A 2 Instrument Data and Operating Conditions 27...

Page 5: ...s are taken based on the results 1 2 Installation and Use The installation procedure shall be carried out only by authorized persons who are familiar with the installation instructions Use only access...

Page 6: ...e of flammable gases and vapors Provide fire extinguishing equipment Transportation Empty the measuring cell and all hoses before you move or lift the instrument To carry the instrument put one hand i...

Page 7: ...era image U View Intuitive user interface Operate the instrument via the touchscreen user interface giving you the look and feel of a smart phone If you prefer you can optionally connect an external k...

Page 8: ...er jack 8 On Off switch 9 Safety temperature limiter A and reset button B 10 Type plate with serial number P N mat no Air connectors on the rear 11 Air pump 4 5 mm barbed inlet C outlet D 12 Drying of...

Page 9: ...nough room approx 10 cm 4 in between the cooling fins on the rear of the instrument and any wall Also ensure that the heat can freely dissipate Do not install the instrument in a cabinet The instrumen...

Page 10: ...f optional parts refer to the particular instructions coming with the part or to the corresponding section in the reference guide 3 2 6 Power Connection 1 Connect the power supply to the DC power jack...

Page 11: ...but during transport the density adjustment may have been compromised If both checks succeed your instrument is ready for routine measurements Else perform an air water adjustment as described in sec...

Page 12: ...Use time server to Yes The Time servers input field replaces the input fields Date and Time c The Time servers setting depends on your network settings If you have set up an automatic network configur...

Page 13: ...details You can place the following widgets onto a dash board of the instrument the live camera image of the measuring cell U View measurement information giving information about the current status...

Page 14: ...Yes is the standard setting Select No if the detection of filling errors by FillingCheck obviously contradicts what you see in the live camera image TIP FillingCheck provides excellent support to the...

Page 15: ...in the menu then select the specific product b Check the measurement settings or edit them then tap Create Measurement To define all measurement settings on the fly and perform a measurement without...

Page 16: ...ll corrode the wetted parts Sample leaking from corroded parts may cause serious injuries Before you fill any sample or cleaning liquid in particular hazardous or flammable chemicals into the instrume...

Page 17: ...is due to the intrinsic properties of glass which the measuring cell is made of If you want to achieve best measurement results give the measuring cell and sample enough time for temperature condition...

Page 18: ...large enough to leave some space for creating the vacuum Only up to two thirds of the syringe should be filled up Boiling the sample 1 Boil the liquid for several minutes to remove dissolved gas 2 Fi...

Page 19: ...lid again Examine the camera image included in the results to check that the water has been filled bubble free Check the quality of the water Clean the measuring cell thoroughly If above actions do no...

Page 20: ...TIP The atmospheric pressure that you get from a local weather station is usually not the absolute atmospheric pressure but a calculated atmospheric pressure value referenced to sea level b Enter the...

Page 21: ...nts See the manual of Xsample 200 High Resistance for a description of the cleaning and drying proce dure with the optional peristaltic pump 1 Select Cleaning in the menu 2 Rinse the measuring cell wi...

Page 22: ...glass surface which are difficult to remove If you store the instrument for less than one day the measuring cell can be filled with ultra pure water WARNING Before using any cleaning agents for the in...

Page 23: ...he instrument are generally excluded from the warranty as well 10 2 Repair Performed by an Autho rized Anton Paar Representative In case your instrument needs repair contact your local Anton Paar repr...

Page 24: ...ideal conditions and for low densities viscosities 0 001 g cm3 0 0001 g cm3 Temperature Measuring range 15 C to 40 C 59 F to 104 F 15 C to 60 C 59 F to 140 F Repeatability s d a 0 1 C 0 2 F 0 02 C 0 0...

Page 25: ...degree 2 Overvoltage category II for the supplied AC adapter Display 7 TFT WVGA 800x480 px PCAP touchscreen Controls touchscreen optional keyboard mouse and barcode reader Interfaces 1x Ethernet 3x US...

Page 26: ...Appendix B EU Declaration of Conformity Appendix B EU Declaration of Conformity...

Reviews: