ICS Schneider Messtechnik Anton Paar DMA 501 Instruction Manual And Safety Information Download Page 19

8 Checks, Adjustments, Calibrations

E15IB001EN-N

19

8 Checks, Adjustments, Calibrations

8.1  Checks

1. Select 

Checks

 in the menu, then select one 

of the already defined checks (or define a new 
one by tapping 

).

2. To edit the check parameters, tap 

.

You can define checks for virtually any measur-
ing quantity available on the instrument.
a. Edit the parameters as appropriate.

For a 

water check

 or 

air check

 select as 

follows:
- target type: “Quantity”
- target quantity: “Water density” or “Air 

Density”, respectively

For a 

custom check

 select:

- target type: “Constant”
and specify a target value to be tested with 
the check.

b. Also specify max. upper and lower deviation 

and the correct temperature for the check.

c. Save the settings.

3. Observe all hints on the screen and fill the 

check medium.
- For a 

water check

, use freshly degassed 

ultra-pure water.

- For an 

air check

, clean and dry the measur-

ing cell thoroughly.

Use the camera image to check that the measur-
ing cell is clean or that water has been filled 
bubble-free.

4. Tap 

START CHECK

.

5. The summary of the check will show the check 

results and a recommendation if the check has 
failed and an adjustment is necessary.

IMPORTANT: 

We advise you to follow the recom-

mendations.

All results are also saved in the data memory.

6. Tap 

DONE

 to close the summary.

If the water check has failed

We recommend taking corrective actions until the 
check is valid again:

• Examine the camera image included in the 

results to check that the water has been filled 
bubble-free.

• Check the quality of the water.
• Clean the measuring cell thoroughly.
• If above actions do not help, perform an 

air/water adjustment.

8.2  Adjustments

With an adjustment, a sample of exactly known 
measurement properties (standard) is measured, 
and the instrument constants are adjusted in a way 
that the instrument delivers the known correct 
results.

The instrument features a 2-point adjustment 
(air/water adjustment) and a wide range adjustment. 
DMA 1001 also features a 1-point adjustment 
(water adjustment) as the standard adjustment.

DMA 1001:

 After ten 1-point adjustments you are 

prompted to perform a 2-point adjustment for 
accuracy reasons.

• Perform a density adjustment if a check has 

failed and corrective actions have not helped.

• We recommend an air/water density adjustment 

if you relocate the instrument to a place where 
the ambient temperature differs by more than 
10 °C (18 °F).

DMA 1001 only:

 To achieve highest accuracy, 

we recommend to perform an air/water density 
adjustment after each temperature change (one 
hour after the required temperature has been 
reached).

• You will likely never need a wide range adjust-

ment. Be aware that it will last several hours. 
However, a wide range adjustment may be 
useful when you adjust your instrument at a 
temperature in the middle of the range, and if 
subsequent measurements are out of specifica-
tions at other temperatures.

Table 2: Factory settings for max. deviations 

(density)

DMA 501:

0.001 g/cm

3

DMA 1001:

0.0002 g/cm

3

Summary of Contents for Anton Paar DMA 501

Page 1: ...Instruction Manual and Safety Information DMA 501 DMA 1001 Density Meter instrument software version from 4 1 0 Original Instructions...

Page 2: ...and its contents may be changed or amended by Anton Paar at any time without prior notice All rights reserved including translation This document or any part of it may not be reproduced changed copie...

Page 3: ...13 6 Measurement Settings Products 14 6 1 Products 14 6 2 Measurement Settings 14 6 2 1 Dashboards 14 6 2 2 Reports 14 6 3 Measurement Parameters 14 6 3 1 Set Temperature 14 6 3 2 Measurement Perform...

Page 4: ...ed Anton Paar Service Engineer 25 10 2 Repair Performed by an Authorized Anton Paar Representative 25 Appendix A Technical Data 26 A 1 Specifications 26 A 2 Instrument Data and Operating Conditions 27...

Page 5: ...s are taken based on the results 1 2 Installation and Use The installation procedure shall be carried out only by authorized persons who are familiar with the installation instructions Use only access...

Page 6: ...e of flammable gases and vapors Provide fire extinguishing equipment Transportation Empty the measuring cell and all hoses before you move or lift the instrument To carry the instrument put one hand i...

Page 7: ...era image U View Intuitive user interface Operate the instrument via the touchscreen user interface giving you the look and feel of a smart phone If you prefer you can optionally connect an external k...

Page 8: ...er jack 8 On Off switch 9 Safety temperature limiter A and reset button B 10 Type plate with serial number P N mat no Air connectors on the rear 11 Air pump 4 5 mm barbed inlet C outlet D 12 Drying of...

Page 9: ...nough room approx 10 cm 4 in between the cooling fins on the rear of the instrument and any wall Also ensure that the heat can freely dissipate Do not install the instrument in a cabinet The instrumen...

Page 10: ...f optional parts refer to the particular instructions coming with the part or to the corresponding section in the reference guide 3 2 6 Power Connection 1 Connect the power supply to the DC power jack...

Page 11: ...but during transport the density adjustment may have been compromised If both checks succeed your instrument is ready for routine measurements Else perform an air water adjustment as described in sec...

Page 12: ...Use time server to Yes The Time servers input field replaces the input fields Date and Time c The Time servers setting depends on your network settings If you have set up an automatic network configur...

Page 13: ...details You can place the following widgets onto a dash board of the instrument the live camera image of the measuring cell U View measurement information giving information about the current status...

Page 14: ...Yes is the standard setting Select No if the detection of filling errors by FillingCheck obviously contradicts what you see in the live camera image TIP FillingCheck provides excellent support to the...

Page 15: ...in the menu then select the specific product b Check the measurement settings or edit them then tap Create Measurement To define all measurement settings on the fly and perform a measurement without...

Page 16: ...ll corrode the wetted parts Sample leaking from corroded parts may cause serious injuries Before you fill any sample or cleaning liquid in particular hazardous or flammable chemicals into the instrume...

Page 17: ...is due to the intrinsic properties of glass which the measuring cell is made of If you want to achieve best measurement results give the measuring cell and sample enough time for temperature condition...

Page 18: ...large enough to leave some space for creating the vacuum Only up to two thirds of the syringe should be filled up Boiling the sample 1 Boil the liquid for several minutes to remove dissolved gas 2 Fi...

Page 19: ...lid again Examine the camera image included in the results to check that the water has been filled bubble free Check the quality of the water Clean the measuring cell thoroughly If above actions do no...

Page 20: ...TIP The atmospheric pressure that you get from a local weather station is usually not the absolute atmospheric pressure but a calculated atmospheric pressure value referenced to sea level b Enter the...

Page 21: ...nts See the manual of Xsample 200 High Resistance for a description of the cleaning and drying proce dure with the optional peristaltic pump 1 Select Cleaning in the menu 2 Rinse the measuring cell wi...

Page 22: ...glass surface which are difficult to remove If you store the instrument for less than one day the measuring cell can be filled with ultra pure water WARNING Before using any cleaning agents for the in...

Page 23: ...he instrument are generally excluded from the warranty as well 10 2 Repair Performed by an Autho rized Anton Paar Representative In case your instrument needs repair contact your local Anton Paar repr...

Page 24: ...ideal conditions and for low densities viscosities 0 001 g cm3 0 0001 g cm3 Temperature Measuring range 15 C to 40 C 59 F to 104 F 15 C to 60 C 59 F to 140 F Repeatability s d a 0 1 C 0 2 F 0 02 C 0 0...

Page 25: ...degree 2 Overvoltage category II for the supplied AC adapter Display 7 TFT WVGA 800x480 px PCAP touchscreen Controls touchscreen optional keyboard mouse and barcode reader Interfaces 1x Ethernet 3x US...

Page 26: ...Appendix B EU Declaration of Conformity Appendix B EU Declaration of Conformity...

Reviews: