User Manual/ Ver. 4.8/ Mar. 2015/ PMH-0014-48/ Page: 110
10.
Calibration
10.1
A/D Calibration
For PEX-1202 and PCI-1202/1800/1802 series card:
Step 1: Apply 0 V to channel 0
Step 2: Apply 4.996 V to channel 1
Step 3: Apply +0.6245 V to channel 2 for PEX-1202
L
and PCI-1202/1800/1802
(L/LU)
Step 4: Apply +4.996 mV to channel 2 for PEX-1202
H
and PCI-1202/1800/1802
(H/HU)
Step 5: Run DEMO19.EXE
Step 6: Adjust VR101 until CAL_0 = 7FF or 800
Step 7: Adjust VR100 until CAL_1 = FFE or FFF
Step 8: Repeat Step6 & Step7 until all OK
Step 9: Adjust VR1 until CAL_2 = FFE or FFF
Step 10: Adjust VR2 until CAL_3 = 000 or 001
For PCI-1602 series card:
Step 1: Apply 0 V to channel 0
Step 2: Apply 4.996 V to channel 1
Step 3: Apply +0.6245 V to channel 2
Step 4: Run DEMO19.EXE
Step 5: Adjust VR3 until channel 0 = 0000 or FFFF
Step 6: Adjust VR2 until channel 1 = 7FFF or 7FFE
Step 7: Repeat Step5 & Step6 until all OK
Step 8: Adjust VR1 until channel 2 = 0FFC or 0FFD