Technical specifications
Model
STE-18-D Pro
STE-23-D Pro
Chamber (mm)
250 x 350
250 x 450
Overall dimensions (mm)
490(W)*455(H)*600(D)
490(W)*455(H)*690(D)
Net Weight (kg)
50
53
Nominal power (VA)
1750
1750
Rated Voltage
220-240V / 110-120V ; 50/60 Hz
Sterilization temperatures
121°C/134°C
Capacity of the distilled
water tank
2.5 L (Water at level Max.) Approx.
0.5L (Water at level Min.) Approx.
Circuit breaker
F16A /400 V
Operation temperature
5°C ~ 40°C
Operation relative humidity
Max. 80%, non-condensing
Max. Noise level
<70 dB
Atmospheric pressure
76 kPa ~ 106 kPa
Packing content
Item
Accessories
Quantity
1
Instruments tray
3
2
Instruments tray
rack
1
3
Draining hose
2
4
Door Seal
1
5
Tray handle
1
6
Door adjustment
wrench
1
7
Instructions Manual
1
3 Installation
General conditions
Position the device on a plane surface with a minimum capacity of 60 kgs.
The sterilizer should be placed on a level worktable.
Leave at least 10 cm between the device rear part and the wall. The clearance required
to open the door is 40cm.
Position the sterilizer at such a height as to make it possible for the operator to check the
whole sterilization chamber and carry out the normal cleaning operations.
The room where the device is installed must be sufficiently ventilated.