24
Appendix 2
Diagrams of the sterilization programs STE-18-D Pro
The time required for sterilizer to be ready for routine use after the power is switched is less than
5 minutes.
The max. Temperature of the 134°C sterilization cycle is 137°C
The max. Temperature of the 121°C sterilization cycle is 124°C
Programs
T
e
m
p
e
ra
tu
re
(
°C
)
P
re
ssu
re
(
kP
a
)
H
o
ld
in
g
t
im
e
(m
in
)
T
o
ta
l t
im
e
(
m
in
)
Type
M
a
x
lo
a
d
(
kg
)
M
a
x
lo
a
d
p
e
r t
ra
y
(kg
)
Solid 134
134
210
3.5
12-30
Unwrapped solid material
4.50
1.20
Wrapped
134
134
210
4
30-49
Unwrapped solid material
4.50
1.20
Single-wrapped solid or hollow
material
4.00
1.10
Wrapped
121
121
110
20
38~54
Unwrapped solid material
4.50
1.20
Single-wrapped solid or hollow
material
4.00
1.10
Prion 134
134
210
18
37-52
Unwrapped porous material
1.00
0.30
Single-wrapped porous
material
0.80
0.25
Dual-wrapped porous material
0.60
0.20
Single-wrapped hollow
material
3.50
1.00
Dual-wrapped solid and hollow
material
1.50
0.50
Drying
(optional)
―
―
―
1-20
―
―
―
B&D test
134
210
3.5
28-35
―
―
―
Helix test
134
210
3.5
28-35
―
―
―
Vacuum test
―
―
―
20-25
―
―
―