Select “Program” from the main menu by
pressing
then
to enter in the menu,
select “Vacuum test”.
After closing the door, press
to start the cycle.
The stage, conditions and the status of the cycle
will appear on the display. The sterilizers will
perform the program automatically. (see appendix
2).
In compliance with EN 13060, the test requires that
the air leakage rate less than or equal to 0.13
kPa/min during 10 minutes.
If leakage rate is not greater 0.13, it will show Success.
If the temperature difference between the max. Temperature and the Min. is above 3°C, it will
show void. That means the result of the test is fail. You need run the vacuum test again after the
chamber has cooled down.
Descale
Mix the descaling powder with distilled water to make descaler according with the
instruction of the descaling powder manufacturer.
From the Programs menu select descale.
After closing the door, press
to start the cycle. The stage, conditions and the status of the
cycle will appear on the display. The sterilizers will perform the program automatically. (see
appendix 2).
Follow the instructions in the screen.
Remove all instruments inside the chamber a nd close the door.
Empty the freshwater tank and fill with descaler which prepared as above, to 2/3 of the water
tank.
Press
, start descaling, it will take around 25 minutes.
After descaling, empty the freshwater tank and fill with distilled water.
Press , start flushing, it will take around 8 minutes.
After flushing, the cycle “Descale” is end.
Data
The internal memory will store the information of the last 9999 cycles.