Print labels (Optional)
Select “labels”, it will show cycle numbers.
Select the cycle number, press button to
print the labels.
6 Maintenance
To ensure proper operation and maximum steam sterilizer life, carefully follow all
recommendations for periodic maintenance.
One of the most important steps you can take to prevent problems with your sterilizer is to use
only distilled water.
Frequency
Number of cycles
Maintenance operation
Monthly
50
Clean the door seal
Clean the filter inside the chamber and in the
clean water tank
Clean the chamber the trays and the rack
Clean the external surface
Every 3 months
200
Clean the distilled water tank
Every year
800
Replace the door seal
Select Settings from the main menu and tap “Maintenance”, to enter in the menu. Select
the item to check the information.
Clean the distilled water tank
Disconnect the main cable.
Drain the tank completely using the drain connector at the back of the sterilizer and leave it
connected into the connector in an open position.
Clean the internal surface with a soft sponge and a small soft brush for the areas difficult to reach
using and a distilled water.
Remove the filter and clean it with a small soft brush and mild soap, rinse it with distilled water,
and put it back in to the position.
Replacement of the bacteriological filter.
The bacteriological filter is in the back of the sterilizer. Unscrew
the filter by hand anti-clockwise.
Place the new bacteriological filter.
Screw the new filter by hand clockwise.
Note: Do not operate sterilizer without filters in place.