ICANCLAVE STE-18-D Instruction Manual Download Page 21

 

21 

 

 

7 Troubleshooting 

 
 

Code 

Description 

Proposed solution 

E1 

Steam generator temperature 
sensor error. 

Power off & run a new cycle  
Contact your supplier if error persists. 

E2 

Inner temperature sensor error. 

Power off & run a new cycle 
Contact your supplier if error persists. 

E3 

Temperature sensor of the 
chamber wall error. 

Carefully ensure that the chamber wall is heated 
and contact your supplier. 

E5 

Fail to release the pressure. 

Power off & run a new cycle 
Contact your supplier if error persists. 

E6 

Door lock problem during the cycle. 

Make sure you had closed the door properly. 
check the door switch. 

E7 

Error between temperature and 
pressure correlation. 

Power off & run a new cycle 
Contact your supplier if error persists. 

E8 

Error between temperature and 
pressure correlation. 

Power off & run a new cycle 
Contact your supplier if error persists. 

E9 

Failure to hold temperature. 

Ensure the distilled tank isn't empty. Check the 
inner temperature sensor. Check somewhere for 
leaking. 

E10 

The door locking system doesn’t 
work. 

The electromagnet of locking system doesn’t 
work. 
The switch of locking system doesn’t work. 

E11 

Failure to preheat the steam 
generator. 

Power off & run a new cycle 
Contact your supplier if error persists. 

E12 

Failure to preheat the chamber. 

Power off & run a new cycle 
Contact your supplier if error persists. 

E13 

Vacuum failed. 

Power off & run a new cycle 
Contact your supplier if error persists. 

E15 

Inner temperature sensor error #2* 

Power off & run a new cycle 
Contact your supplier if error persists. 

E16 

Pressure error  

Replace the air filter 
Power off & run a new cycle 
Contact your supplier if error persists. 

N20 

Program manually interrupted 

Reset the error from main screen. 

E22 

Vacuum test failure 

Somewhere is leaking. Check the door seal. 
 Or contact your supplier if error persists. 

N23 

Result of vacuum test is void 

The temperature of the chamber is high. 
Try again after the chamber has cooled down. 

E24 

It takes too long time to enter the 
next status. 

Check somewhere leaking.  
Or contact your supplier if error persists. 

N27 

The vacuum test fails. 

Switch off. Then switch on after the chamber cool 
down and try again. 

 

1. Press the included 

hose on to the drain valve 

firmly. 

 

2.  Turn the drain valve 

clockwise to drain the 

tank. 

 

3.  Push and turn the drain 

valve anticlockwise to be 

back after draining the tank. 

Summary of Contents for STE-18-D

Page 1: ...Instructions Manual For Models STE 18 D Pro STE 23 D Pro...

Page 2: ...fully and follow all installation instructions Ningbo Ican Machines Co Ltd No 77 Yunlin East Road Gulin Town Ningbo China www icanclave com European Representative Icanclave Europe S L Juan Ramon Jime...

Page 3: ...3...

Page 4: ...ect the program 13 Start the sterilization program 14 End of cycle 14 Manually interruption of the cycle 15 Delay start cycle 15 Test Program 16 Descale 16 Data 17 Print and save report Optional 18 Pr...

Page 5: ...only to a properly grounded outlet Do not cover or block any openings on this appliance Use this appliance only for its intended use as described in this manual Do not exceed the maximum weight limit...

Page 6: ...door is open Fill the tank if it is flashing Min level of freshwater tank 0 5Liters Drain the tank if it is flashing The capacity is 1 5 liters 1 Report printer Grey No connected no paper Green Ready...

Page 7: ...vent 3 Door lock 11 USB port 19 Condenser vent 4 Door handle 12 Main switch 20 Bacterial filter 5 Chamber 13 Distilled water tank vent 21 Used water tank vent optional 6 Door interlock 14 Used water t...

Page 8: ...essure 76 kPa 106 kPa Packing content Item Accessories Quantity 1 Instruments tray 3 2 Instruments tray rack 1 3 Draining hose 2 4 Door Seal 1 5 Tray handle 1 6 Door adjustment wrench 1 7 Instructions...

Page 9: ...s recommended Never connect the device pin to reductions of any type Location requirements mm 4 Setup Switch On and open the door to remove all of the inner contents for unpacking Connect the power co...

Page 10: ...ument manufacturer s guidelines and recommendations for handing and cleaning instruments prior to sterilization Check the manufacturer s instructions as to proper procedure for sterilizing of each ite...

Page 11: ...The Setup menu permits to set the following options Parameter Unit Preheat Expire date labels Water quality sensor Preheat Factory reset and see the information of Last error Select Setup from the mai...

Page 12: ...um temperature to begin a sterilization cycle this helps to reduce the total cycle time and the drying efficiency The Preheat mode will be deactivated after one hour of inactivity Select Preheat from...

Page 13: ...st it by pressing Press button to save and back Last Error In order to help the technical assistance process the most relevant information corresponding to the last error can be displayed on the scree...

Page 14: ...n Select Program Select the program by pressing then press to confirm program in the screen will appear the information of selected program as the temperature and sterilization time holding time also...

Page 15: ...Note If the cycle is manually interrupted before it reaches the drying phase the items inside the sterilizer must be considered not sterile N20 will appear on the screen see Error code description Ca...

Page 16: ...perform the program automatically see appendix 2 After finishing the cycle you may check the indicator and evaluate the result according with the instructions of the test manufacturer B D test Put th...

Page 17: ...ooled down Descale Mix the descaling powder with distilled water to make descaler according with the instruction of the descaling powder manufacturer From the Programs menu select descale After closin...

Page 18: ...four numbers represent cycle number The last two numbers represent error code 00 no error 01 error E01 Sample of a printer report Program WRAPPED Temperature 134C Pressure 206 0 kPa Drying Time 08Min...

Page 19: ...oor seal Select Settings from the main menu and tap Maintenance to enter in the menu Select the item to check the information Clean the distilled water tank Disconnect the main cable Drain the tank co...

Page 20: ...you may adjust it Open the door insert your finger into the bottom of the door and pull the ring to unlock the mechanism Turn the lid counterclockwise to tighten it If it is too tight you may also tur...

Page 21: ...t work The switch of locking system doesn t work E11 Failure to preheat the steam generator Power off run a new cycle Contact your supplier if error persists E12 Failure to preheat the chamber Power o...

Page 22: ...tion Temporary interruption of winding 3 Safety valve Protection against possible sterilization chamber over pressure Action Release of the steam and restoration of the safety pressure 4 Safety micro...

Page 23: ...oxide sio2 1mg I 1 0mg kg Iron 0 2mg I 0 1mg kg Cadmiun 0 005mg I 0 05mg kg Lead 0 05mg I 0 1mg kg Rest of heavy metals 0 1mg I 0 1mg kg Chloride 2mg I 0 1mg kg Phosphates 0 5 mg I 0 1mg kg Conductivi...

Page 24: ...134 134 210 3 5 12 30 Unwrapped solid material 4 50 1 20 Wrapped 134 134 210 4 30 49 Unwrapped solid material 4 50 1 20 Single wrapped solid or hollow material 4 00 1 10 Wrapped 121 121 110 20 38 54...

Page 25: ...id 134 134 210 4 15 30 Unwrapped solid material 5 00 1 50 Wrapped 134 134 210 4 38 57 Unwrapped solid material 5 00 1 50 Single wrapped solid or hollow material 4 50 1 20 Wrapped 121 121 110 20 46 62...

Reviews: