Manufacturer’s guidelines and declaration – electromagnetic immunity
SONOPEEL is destined to be used in the electromagnetic environment specified
below. The user of the equipment should ensure that it will be used in such an
environment.
Immunity test
Test level
IEC 60601
Conformity level
Electromagnetic
environment - guidelines
Electrostatic
discharge
(ESD)
IEC 61000-4-2
6 kV by
contact
8 kV by air
6 kV by contact
8 kV by air
Floors should be wooden,
concrete or ceramic. If
floors are covered with
synthetic
material,
the
relative humidity should
be at least 30%.
Fast electric
transitories /
pulse train
(Burst)
IEC 61000-4-4
2 kV in the
feeding lines
1 kV in the
input/output
lines
2 kV in the
feeding lines
1 kV in the
input/output lines
The quality of power
supply should be that of a
typical
hospital
environment
or
a
commercial building.
Outbreaks
IEC 61000-4-5
1 kV
differential
mode
2 kV common
mode
1 kV differential
mode
2 kV common
mode
The quality of power
supply should be that of a
typical
hospital
environment
or
a
commercial building.