Adjustment and Calibration
Flow measurement
This mode is available only if a flow sensor is connected.
→
Flow
(available only in version
BL and BQ,
if enabled in the Basic Settings)
→
Attenuation
→
reading min. value
→
reading max. value
Returns automatically back to measuring mode after 5 sec
→
reset of both the min. and max. value
→
return to measuring display
→
Reset accumulated volume to 0
The flow sensor must be connected with
the interface shown in the picture below.
Sensor Out Filter Sensor In
The head line of the display shows the measuring unit
ml/min
and after a space the unit for the accumulated
value as
l
for liter. In the middle line the flow in ml/min and in the bottom line the accumulated value in liter
is shown.
For high accuracy measurement we recommend to rinse the flow sensor with 100 ml/min RO water for 15
min.
Reset accumulated value
The accumulated flow value can be resettled at any time following the procedure below.
Procedure
1.
Press and hold down the key
until the upper display line shows "ZERO", this will take
three seconds
3. The accumulated value is now set to zero.
Attenuation of reading
The reading of the current can be attenuated.
With button the selection menu can be called.
Values up to 180 can be selected. The selected value represents the time in seconds, during the values are
averaged.
To avoid malfunction of the sensor, please make
sure to connect fluid correctly the sensor. The
sensor has a label showing the flow direction.