19
Adjustment and Calibration
Temperature compensation
Listed below are the temperature coefficients used by most major hemodialysis machine and reverse osmosis
manufacturers. We recommend that you double check this data with documentation from your machine’s
manufacturer
Display
To use for
Temperature
coefficient
at 25°C
Temperature
coefficient
at 37°C
BAXTER
Baxter (Europe Dialysis-Machine)
2,200 %/K
2,200 %/K
BELLCO
Bellco
2,000 %/K
2,100 %/K
B.BRAUN
B.Braun
2,000 %/K
2,100 %/K
FRESENI
Fresenius
2,000 %/K
2,100 %/K
GAMBRO
Gambro
AK-Type and Phoenix Machines
2,000 %/K
2,070 %/K
HOSPAL
Hospal
2,000 %/K
2,070 %/K
NIK ALL
Nikkiso Overall Conductivity
2,050 %/K
2,050 %/K
NIK BIC
Nikkiso Bicarbonate Conductivity
2,200 %/K
2,200 %/K
STANDARD Other Dialysis Machines
2,000 %/K
2,070 %/K
USER TC
Adjustable
0 … 5,000 %/K
0 … 5,000 %/K
nLF-ISO
Non Linear Function for
tap, purified and waste water.
According Standard
ISO7888
– DIN EN27888
Reference temperature 25°C
RO-DWA
Reverse Osmosis DWA
DWA GmbH & Co. KG
Ubstadter Straße 28
76698 Ubstadt-Weiher
0,00 %/K
RO-HERCO Reverse Osmosis Herco
Herco Wassertechnik GmbH
Planckstraße 26
71691 Freiberg
2,200 %/K
Reference temperature 25°C
RO-LAUER Reverse Osmosis Lauer
LAUER MEMBRAN WASSERTECHNIK GmbH
Industriegebiet Speichermatt
79599 Wittlingen
nLF-ISO according Standard
ISO7888
– DIN EN27888
Reference temperature 25°C
RO-PPW
Reverse Osmosis PPW
Phoenix Pure Water GmbH
Siemens Technopark
Werner-von-Siemens-Str. 2-6
D-76646 Bruchsal - Germany
2,200 %/K
Reference temperature 25°C