ibidi Heating System, Universal Fit
NOTE
The temperature of the Heated Lid must be set to at least some degrees warmer than the temper-
ature of the Heated Plate!
The temperature adjustment must be repeated from time to time (at least once a year), especially if
one of the following conditions has been changed:
• Room temperature
• Air conditioning
• Chamber type or objective lens
• Humidification and gas flow
• Use of an XL-Incubator
• Use of an Objective Heater
5.5
Sample Preparation
Check that the cell cultureware you intend to use fits into the Insert of the ibidi Heating System,
Universal Fit.
Prepare the cells according to your protocol and place the culture vessel in the Heating Insert. The
Heating Insert can be removed from the ibidi Heating System, Universal Fit even with the holder in
place. Then put the whole assembly in the opening of the Heated Plate and close the Heated Lid.
NOTE
Let the temperature of the system equilibrate for a minimum of 30 min before you start your
experiments.