ibidi Heating System, Universal Fit
1.3
Nomenclature
Temperature Controller
Heated Lid
Heated Plate
1.4
Specifications
Only operate the Heating System in the range of the specifications given below:
Table 1 – Specifications of the ibidi Heating System, Universal Fit
Electrical Specifications Power Supply
Protection class
I
International protection marking
(IEC 60529)
IP 20
Overvoltage category
II
External power supply
AC 100-240 V, 50/60 Hz, 2 A
Input line voltage Temperature
Controller
DC 24 V, 6.67 A, 160 W
Output
voltage
to
channel
1
(Heated Lid), Universal Fit
DC 10 V, max. 5 A
Output
voltage
to
channel
2
(Heated Plate), Universal Fit, for 1
Chamber
DC 12 V, max. 5 A
Operating and Storage Conditions
Operating site
Indoor use only
Operating temperature
18-30°C/64-86°F (min 5°C/9°F less than set temperature)
Operating humidity
max. 80% relative humidity (RH)
Operating altitude
max. 2000 m (atmospheric pressure 800-1060 hPa/11.6-15.4
psi)
Storage conditions
-5-50°C/23-122°F, humidity
<
60% relative humidity (RH)
Outer Dimensions and Characteristics of the Components
Temperature Controller
90 mm × 170 mm × 230 mm
Weight: 1720 g/3.8 lbs
Heated Lid
19 mm × 85.5 mm × 127.5 mm (134.5 mm with cover ridge)
Length of cable: 1.5 m
Instruction Manual
7
Version 2.2 (May 23, 2019)