Manual ibaPADU-8-M
24
Issue 1.9
10.1.4
Input Resources iba FOB-M/IN
The following table shows the input resources of the first link on the first ibaFOB-card.
Up to four ibaFOB-4i cards with four links each are possible.
For further information please refer to the ibaLogic manual.
The device provides actual measured values and ac-
tual parameter settings in return as well.
Number of the activated ibaPADU-8-M (0..95)
Sample time of the activated device in
s
\
|
| Actual gain of the channels
| in dB for channels 1..8
|
These parameters (...Gainx) are always 0!
I (no gain)
|
|
/
\
|
|
| Actual corner frequency of low-pass filter
| for channels 1..8
|
These parameters (...Frequencyx) are always 0!
I (no lowpass filter)
|
/
Actual reset status
Actual command being processed
\
|
|
|
| Measured values of analog channels 1..8
| (signed integers)
|
|
/
\
|
|
|
| Measured values of binary channels 1…8
|
|
/
Status of data buffer (true, if number of values > buffer size)
Quantity of available data (multiples of 10), if data available.
Link status (true, if iba FOB M active and link available)
True, if link status ok and ibaPADU-8-M activated.
True, if data reading speed was too slow.
True, if data overflow. Data acquisition will be stopped!
True, if data acquisition is running.