77
10.3.15 Aggregation submodule
The Aggregation submodule is a freely configurable module in which the measurement
interval and the characteristic values can be selected individually. The submodule name
is assigned automatically by ibaPDA and is in accordance with the set measurement
interval. The default setting is 10 min. If the measurement interval is modified, the module
name will change accordingly.
Note
Details on the aggregation method:
For the standard update intervals (200 ms, 3 s, 10 s, 10 min, 2 h) the aggre-
gation method is listed in the
table „Calculated characteristic values“ in chap-
ter 8.2.
The following aggregation is applied for the "Custom" update interval:
Energy value: total of 10/12 period values
Flicker: Cubic average of Pst values
For all other calculations the quadratic average is being used.
“General” tab
Basic settings
See Power frequency submodule, “General” tab, chapter 10.3.2
Module Layout
No. analog signals
Enter the number of desired signals here. The number determines the length of the signal
table in the “Analog” tab.
Configuration
Update interval
Select the measurement interval here.
Summary of Contents for ibaPQU-S
Page 6: ...Manual ibaPQU S 6 Issue 1 5 14 Index 125 15 Certificate 126 16 Support and contact 127...
Page 23: ...Manual Issue 1 5 23...
Page 90: ...Manual ibaPQU S 90 Issue 1 5 Distance between two ibaPQU S systems Dimensions in mm...
Page 91: ...Manual Issue 1 5 91 ibaPQU S and backplane Dimensions in mm...
Page 92: ...Manual ibaPQU S 92 Issue 1 5 Dimensions of ibaPADU S B1S with modules dimensions in mm...
Page 106: ...Manual ibaPQU S 106 Issue 1 5 12 3 2 3 Dimensions Dimensions in mm...
Page 126: ...Manual ibaPQU S 126 Issue 1 5 15 Certificate...