IACER Srl
6
MNPG148-06
MIO-CARE PRO is used to apply electrical micro impulses which create energy.
This energy, modulated with different parameters specific for different
impulses, can give the patient many benefits from pain relief to muscles cool
down, from muscles strengthening to drainage, from isotonic exercises to
hematomas’ treatment, to the treatment of the imperfections using the beauty
programs.
The patient population intended for electrotherapy treatment using the MIO-
CARE PRO device includes patients of both sexes, men and women, of age
(unless otherwise indicated by medical doctors). For further details, please
refer to the Contraindications section.
The CE0068 mark is only for the medical programs (see the following
paragraphs related to the detailed description of the programs).
Caratteristiche tecniche
Characteristics
Specifications
Power supply
Rechargeable batteries AAA Ni-MH 4.8V
800mAh
Recharger
power supply line AC 100-240V, 50/60Hz
200mA;
Output DC 6.8V, 300mA max.*
Isolation (EN 60601-1)
II
Applied part (EN 60601-1) BF
Protection level
IP22
Applied part to the patient Electrodes
Dimensions
(length x height x depth)
260x176x60mm
Weight main body
205gr including batteries
Layout
ABS
Number output channel
2 independent
Functioning
Continuous
Intensity
adjustable
Max output current
99mA, 1K
Ω
load each channel in the
remaining programs
Impulse
Biphasic compensated square wave and
monophasic square wave
Frequency
From 0.25 to 200Hz
Impulses’s width
From 20 to 450
µ
s
Therapy
Time depending on the program (1-90 min)
Display
Reflective and illuminated LCD display