Session duration: 24 minutes.  

Electrodes’ positioning:  photo from 01 to 23 (21 excluded) of the 

Positions 

manual

Intensity: gradually increased session after session without overexerting the 

muscles until reaching the maximum level of tolerance. 

 

NEMS10 • Explosive strength upper limbs and trunks (non-medical program) 

NEMS11 • Explosive strength lower limbs (non-medical program) 

Explosive strength programs increase the explosive power and speed of the 

muscle mass, with extremely short, strengthening contractions and very long 

active recovery times to allow the muscle to regain strength. It is usually best 

to complete a three-week basic strength cycle of increasing intensity, before 

using this program. Then continue with three weeks of explosive strength twice 

a week. 

Session duration: 24 minutes.  

Electrodes’ positioning:  photo from 01 to 23 (21 excluded) of the 

Positions 

manual

Intensity: during contraction, the intensity has to be the highest that can be 

endured in order to obtain maximum muscle exertion whilst involving the 

greatest number of fibres. 

 

NEMS12 • Deep capillarization (non-medical program) 

This program significantly increases arterial flow in the treated area. Prolonged 

use of this program develops the intramuscular capillary network of fast-twitch 

fibers. The effect obtained is an increase in the capacity of fast-twitch fibers to 

withstand strain over extended periods of time. For an athlete with good 

resistance, the capillarization program is very useful for recovery after intense 

aerobic work, before anaerobic work and when training is not possible (due to 

bad weather or an injury).  

Session duration: 30 minutes. 

Electrodes’ positioning:  photo from 01 to 23 (21 excluded) of the 

Positions 

manual

Intensity: medium. 

 

NEMS13 • Muscle recovery (non-medical program) 

Can be used for all sports, after competitions or the most demanding training 

sessions, in particular after long and intense exertion – to be used immediately 

after exertion.  This program helps drainage and winding down, improving 

muscle oxygenation and helping to discharge synthetic substances produced 

during exertion.  

Summary of Contents for MIO CARE PRO

Page 1: ...USER MANUAL Electrotherapy MIO CARE PRO...

Page 2: ......

Page 3: ...WARNING 11 PATIENT PREPARATION 14 DEVICE USE 14 Operating instructions 15 TENS programs 16 Treatment programs for TENS therapy 24 BEAUTY programs 26 Treatment programs for muscles firming up and lipo...

Page 4: ...and manufactured according to the European Medical Device Directive 93 4 EEC transposed in Italy by the D Lgs 46 97 as modified by the Directive 2007 47 EC D Lgs 37 2010 and further modifications inte...

Page 5: ...ic Environmental intended use Ambulatory and home MIO CARE PRO is indicated for the treatment and the functional rehabilitation of the following pathologies and anatomical zones wrist articulation han...

Page 6: ...the medical programs see the following paragraphs related to the detailed description of the programs Caratteristiche tecniche Characteristics Specifications Power supply Rechargeable batteries AAA N...

Page 7: ...ive humidity From 10 to 90 Atmospheric pressure From 700 to 1060hPa WARNING the device has an output current over 10mA Use only the battery recharger given by the manufacturer The use of other recharg...

Page 8: ...H2 10 Increase program 11 Decrease program 12 ON OFF and OK button 13 Set programs and therapy pause button 14 Belt clip 15 Battery compartment 1 Mode operation NEMS TENS BEAUTY MEM 2 Wave frequency 3...

Page 9: ...on Manufacturer Manufacturing date YYYY MM Read instructions for use The product must be disposed as electronic waste in accordance to WEEE Directive on waste electrical and electronic equipment IP22...

Page 10: ...PRO pack contains n 1 MIO CARE PRO device n 2 connection cables for the transmission of electric impulses n 4 splitting leads n 1 packages containing 4 pre gelled self adhesive 41x41 mm electrodes or...

Page 11: ...en to use MIO CARE PRO if the patient has a pacemaker is cardiopathic suffers from epilepsy is a pregnant woman is an anxious person has severe disease tuberculosis juvenile diabetes viral diseases in...

Page 12: ...s and fire It is forbidden to use the device in the presence of patient monitoring equipment of electrosurgical possible bruises and burns or shortwave or microwave therapy equipment or other equipmen...

Page 13: ...ressure to use pointed or sharp objects on the device keyboard Warning insufficiently sized electrode sections can cause skin reactions or burns do not use damaged electrodes even if they well adhere...

Page 14: ...connect the electrostimulation cable jacks to the self adhesive electrodes position the self adhesive electrodes on the skin see photos of electrode positions in the Positions manual connect the impul...

Page 15: ...tic signal advises the user when the treatment is completed 5 Turn off the device keeping pressed the button for at least two seconds FREE MEMORIES ADJUSTABLE PROGRAMS With MIO CARE PRO you can set th...

Page 16: ...ories according to the mechanism used to reduce the pain The main types are conventional TENS or fast analgesic training TENS or delayed analgesic which is similar to the effect of the electro acupunc...

Page 17: ...ch the muscles surrounding the treated area begin to contract over this limit the stimulation does not become more effective just more irritating so it is best to stop before that point Programs speci...

Page 18: ...s Backache sciatic pain Total time 20 min frequency 90 Hz impulse width 50 s Total time 20 min frequency 60 Hz impulse width 60 s 7 Yes Sprains bruises Total time 10 min frequency 110 Hz impulse width...

Page 19: ...width 150 s Total time 10 min frequency 90 Hz impulse width 50 s 12 Yes Epicondylitis Total time 20 min frequency 90 Hz impulse width 50 s Total time 10 min frequency 70 Hz impulse width 60 s Total t...

Page 20: ...ce an addictive effect meaning that the impulse will be felt less and less if necessary the intensity can be increased by one level to counter this effect Session duration 40 minutes no less than 30 4...

Page 21: ...ction of the muscles surrounding the area treated TENS4 Anti inflammatory medical program Program recommended for inflammatory conditions To be applied until the inflammatory state is lessened 10 15 a...

Page 22: ...Positions manual Intensity to be adjusted between the threshold of perception and pain TENS7 Sprains Bruises medical program The program develops its effectiveness after this type of injury by inhibit...

Page 23: ...s shown in Figure 1 Intensity to be adjusted between the threshold of perception and pain without causing muscle contraction TENS11 Plantar stimulation medical program This program has a relaxing and...

Page 24: ...on affecting the fibrous tissues surrounding joints tendons serous sacs and connective tissue These appear altered and can break into fragments and calcify If neglected this condition can become heavi...

Page 25: ...ENS 3 5 7 Daily up to 2 3 times a day Photo 32 Carpal tunnel syndrome TENS 1 10 12 Daily even twice a day Photo 33 Trigeminal neuralgia TENS 1 10 12 Daily Photo 24 Wryneck TENS 1 TENS 9 8 10 Daily eve...

Page 26: ...x 7sec 80 1Hz 300 s x 3 sec 100 20Hz 300 s x 5 sec 80 x 60 cycles Total time 10 min 3Hz 300 s x 7sec 80 1Hz 300 s x 3 sec 100 30Hz 300 s x 5 sec 80 x 40 cycles 3 No Toning up upper limbs and trunk Tot...

Page 27: ...tal time 10 min 3Hz 300 s x 7sec 80 1Hz 300 s x 3 sec 100 60Hz 300 s x 5 sec 75 x 40 cycles Total time 5 min 3Hz 300 s x 7sec 80 1Hz 300 s x 3 sec 100 70Hz 300 s x 5 sec 75 x 20 cycles 7 No Modelling...

Page 28: ...glutei and hips Total time 4 min frequency 6 Hz impulse width 250 s Total time 20 min 5Hz 250 s x 8 sec ch1 ch2 80 40Hz 250 s x 6 sec ch1 80 40Hz 250 s x 6 sec ch2 80 x 60 cycles Total time 5 min freq...

Page 29: ...g up lower limbs non medical program These programs are indicated for firming up muscles of the arms and bust BEAUTY1 or the legs BEAUTY2 working mainly on slow twitch fibres It is recommended to thos...

Page 30: ...Y4 working mainly on fast twitch fibres It is recommended to those who already practice moderate physical activity Method of use 1 identify the muscle to be treated To obtain good results it is best t...

Page 31: ...ensity should be gradually increased so that muscle contractions are not painful 5 during contraction generated by the unit contract the muscle voluntarily A cycle of 15 20 applications must be comple...

Page 32: ...2 Lipolysis abdomen 9 thighs 10 glutei and hips 11 arms 12 non medical program These specific drainage programs increase microcirculation within and around the muscle fibres treated and create rhythmi...

Page 33: ...BEAUTY15 Heaviness in legs non medical program This program is used to improve blood flow and muscle oxygenation speeding up the elimination of lactic acid produced after anaerobic sessions for muscle...

Page 34: ...UTY1 BEAUTY1 5 Lipolysis abdomen Photo 20 BEAUTY6 BEAUTY 11 BEAUTY6 BEAUTY1 6 Lipolysis thighs Photo 21 BEAUTY7 BEAUTY 11 BEAUTY7 BEAUTY1 6 Lipolysis glutei Photo 19 BEAUTY8 BEAUTY 11 BEAUTY8 BEAUTY1...

Page 35: ...1Hz 200 s x 3 sec 100 30Hz 200 s x 8 sec 80 x 45 cycles 3 No Resistance lower limbs Total time 4 min Frequency 6 Hz impulse width 300 s Total time 15 min 3Hz 300 s x 9sec 80 1Hz 300 s x 3 sec 100 20Hz...

Page 36: ...time 4 min Frequency 6 Hz impulse width 300 s Total time 10 min 3Hz 300 s x 7sec 80 1Hz 300 s x 3 sec 100 50Hz 300 s x 5 sec 75 x 40 cycles Total time 10 min 3Hz 300 s x 7sec 80 1Hz 300 s x 3 sec 100...

Page 37: ...min Frequency 6 Hz impulse width 300 s Total time 10 min 3Hz 300 s x 12sec 90 1Hz 300 s x 3 sec 100 100Hz 300 s x 5 sec 80 x 30 cycles Total time 10 min 3Hz 300 s x 12sec 90 1Hz 300 s x 3 sec 100 120H...

Page 38: ...l tonic contractions lower limbs Total time 3 min Frequency 6 Hz impulse width 300 s Total time 10 min 30Hz 300 s x 5 sec 80 CH1 30Hz 300 s x 5 sec 80 CH2 x 60 cycles Total time 5 min Frequency 4 Hz i...

Page 39: ...0 Total time 10 min Frequency 2 Hz impulse width 250 s 20 No Deep massage Total time 5 min Frequency 3 Hz impulse width 250 s Total time 10 min 3Hz 250 s x 2 sec ch1 100 3Hz 250 s x 2 sec ch2 100 x 15...

Page 40: ...useful for sports involving maximum effort right from the start Suitable for all muscle groups Session duration 16 minutes Electrodes positioning photo from 01 to 23 21 excluded of the Positions manua...

Page 41: ...ntractions alternate with periods of active recovery during the work phase allowing the muscle to be trained without subjecting it to stress and improving oxygenation of the same muscle The following...

Page 42: ...ighest that can be endured in order to obtain maximum muscle exertion whilst involving the greatest number of fibres NEMS12 Deep capillarization non medical program This program significantly increase...

Page 43: ...n duration 24 minutes Electrodes positioning photo from 02 to 05 and 11 12 of the Positions manual Intensity during contraction has to be adjusted so that the contraction intensity is the same as a vo...

Page 44: ...ding training sessions in particular after long and intense exertion to be used immediately after exertion This program helps drainage and capillarization improving muscle oxygenation and helping to d...

Page 45: ...y training program No of weeks Day 1 Day 3 Day 5 Day 7 Abdominal muscles basic strength Photo 1 20 FITNESS6 FITNESS19 FITNESS6 FITNESS6 FITNESS12 5 Pectoral muscles basic strength Photo 7 17 FITNESS6...

Page 46: ...1 90 min frequency 1 200 Hz contraction time 1 10 sec slope 0 5 sec recovery time 0 30 sec width impulse 50 450 s 13 No Battery test M1 M5 TENS Free memories medical program Free memories for antalgi...

Page 47: ...leanliness of the device Device not subject to sterilization Note Never use solvents for cleaning Cleaning agents cause damage to the device Attention to the need for periodic maintenance especially i...

Page 48: ...turer Here below are some typical situations MIO CARE PRO cannot be turned on and or the display does not light up Check the battery status and replace it if it is necessary make reference to chapter...

Page 49: ...into the power socket The display will show the battery blinking icon and the TIME min icon which takes account of the charging time on the display After 4 hours the recharge automatically finishes an...

Page 50: ...vice is not used Discharge the battery as much as possible before the recharging Use only the original battery charger or in any case the battery charger supplied by the fabricant distributor Not open...

Page 51: ...ed to intervene promptly if necessary as a result of manufacturing defects The warranty conditions are those described in the following paragraph Warranty conditions The warranty is provided by IACER...

Page 52: ...ce For any technical assistance contact I A C E R S r l Via Enzo Ferrari 2 30037 Scorz VE Tel 041 5401356 Fax 041 5402684 Technical documentation related to repairable parts could be attached but only...

Page 53: ...careful when applying therapy to patients with pacemakers In any case it is recommended to use the equipment at least at 3 meters away from televisions monitors cellphones or any other electronic equi...

Page 54: ...IACER Srl 54 MNPG148 06 For more details please see the EMC tables at the end of this manual...

Page 55: ...ectromagnetic environment guidance RF emissions CISPR 11 Group 1 MIO CARE PRO uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any inte...

Page 56: ...hat of a typical commercial or hospital environment 1kV for input output lines 1kV for input output lines Impluses IEC 61000 4 5 0 5kV 1kV Line s to line s 0 5kV 1kV Line s to line s Mains power quali...

Page 57: ...onment specified below The customer or the user of the device should assure that it is used in such an environment Immunity test Test level EN 60601 Conformity level Electromagnetic environment guide...

Page 58: ...apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and people a Field strengths from fixed RF transmitters such as base stations for...

Page 59: ...itter m from 150kHz to 800 MHz from 80MHz to 800 MHz from 800MHz to 2 7GHz 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 12 23 For transmitters rated at a maximum output p...

Page 60: ......

Reviews: