System accuracy results for glucose concentrations between
28.6 mg/dL (1.6 mmol/L) and 512 mg/dL (28.4 mmol/L)
Within ±15 mg/dL (±0.83 mmol/L) and Within ±15 %
600/600 (100.0 %)
Precision:
The precision studies were performed in a laboratory
using CareSens N Premier Blood Glucose Monitoring Systems.
Within Run Precision
Blood
average
42 mg/dL
(2.3 mmol/L)
SD = 1.8 mg/dL
(0.1 mmol/L)
72 mg/dL
(4.0 mmol/L)
SD = 2.6 mg/dL
(0.1 mmol/L)
128 mg/dL
(7.1 mmol/L)
CV = 3.9 %
188 mg/dL
(10.4 mmol/L)
CV = 3.1 %
291 mg/dL
(16.2 mmol/L)
CV = 2.9 %