Applying Blood Sample
Step 8
Obtain a blood sample using the lancing
device. Place the device against the pad
of the finger. The best puncture sites are
on the middle or ring fingers. Press the
release button. Remove the device from
the finger. Wait a few seconds for a
3
4
5
6
7
blood drop to form. A minimum volume of 0.5 microliter is
needed to fill the confirmation window (actual size of 0.5 μL: ).
Step 9
After the
symbol appears on the screen, apply the blood
sample to the narrow end of the test strip till the meter beeps. If
the confirmation window is not filled in time because of abnormal
viscosity (thickness and stickiness) or insufficient volume, the Er4
message may appear.
It is recommended to place the test strip vertically into the blood
sample site as shown below.
Good
Sample
Insufficient
Sample