Understanding Error Messages
Maintenance
A used test strip was inserted.
>
Repeat the test with a new test strip.
The blood or control solution sample was
applied before the
symbol appeared.
>
Repeat the test with a new test strip and
wait until the
symbol appears
before applying the blood or control
solution sample.
The temperature during the test was above or
below the operating range.
>
Move to an area where the temperature is
within the operating range (5–50 °C) and
repeat the test after the meter and test
strips have reached a temperature within
the operating range.
The blood sample has abnormally high
viscosity or insufficient volume.
>
Repeat the test after inserting a new test
strip.