Applying Blood Sample
Obtain a blood sample using the lancing
device. Place the device against the pad of
the finger. The best puncture sites are on the
middle or ring fingers. Press the release
button. Remove the device from the finger.
Wait a few seconds for a blood drop to form.
A minimum volume of 0.5 microliter is needed to fill the
Confirmation window. (actual size of 0.5
㎕
:
•
)
After the symbol appears on the screen,
Apply the blood sample to the narrow end of
the test strip till the meter beeps.
If the confirmation window filled before the meter
Finishes down then discard the test strip new one.
If confirmation window is not filled in because of
Abnormal viscosity or insufficient volume,
Er4 message will appear.
Step 8
Step 9
Note:
The meter may switch off if the blood sample is not applied within 2 minutes of
The icon appearing on the screen. If the turns off, remove the strip and
reinsert into the meter.