www.i-sens.com 25
Control Solution Testing
Insert a test strip into the meter’s test strip port, with the contact
bars facing upwards.
Gently push the test strip into the port until the meter beeps.
Be careful not to break the strip while pushing it in.
The icon will be displayed on the screen.
Shake the AutoChek t Control
Solution vial before each test.
Remove the cap and squeeze the
vial to discard the first drop. Then
wipe the tip with a clean tissue
or cloth. After the symbol
appears on the display, apply the
solution to the narrow edge of the
test strip until the meter beeps.
Make sure the confirmation window fills completely.
Caution:
If all the results you get from testing continuously (at least three times)are within the
range printed on the test strip, the meter and test strips are working properly and you
may use them for your blood glucose test.
Step 1
Step 2