11.6.3 Calibration of the probe slope
»
In the main menu, select
2B SLOPE
CALIB
»
Enter the conductivity of the buffer
solution in the "BUFFER" field
according to the table on the bottle
label.
»
Hold the sensor in the buffer solution
and wait until the "READ" value has
stabilized.
»
press
, to confirm the displayed
slope value and exit the parameter
and press
, to return to the main
menu
or
»
press
to exit the parameter with-
out saving
The buffer value must not be 0, otherwise
the calibration is not saved.
11.6.4 Permeate set point calibration
With the assistance of the permeate conduc-
tivity measurement and a freely adjustable
limit value for the permeate conductivity, the
unit can be stopped and an alarm message
issued if the permeate quality is reduced.
»
Select
3 LIMIT OUT
in the main
menu
Press the arrow keys
to select
whether the set point function should be
active or inactive.
»
Confirm the selection with
If the set point function has been activated,
the program jumps to submenu "3B"..
By pressing the arrow keys
you can
set the max. set point
»
Confirm the selection with
. The
program jumps to the submenu "3C".
If the permeate conductivity exceeds the limit
value set here during production, the system
stops, the alarm message "ALARM CON-
DUCTIVITY" appears on the display and the
alarm output relay picks up.
2A ZERO CALIB
2B SLOPE CALIB
READING 12,5
µS
BUFFER 15,0
µS
3 LIMIT OUT
4 LIMIT RAW/MIX
3A LIMIT OUT
ENABLED
3B LIMIT OUT
50,00
µS
Summary of Contents for WL-ROL 160-1200 Series
Page 1: ...FlexLine Reverse osmosis plants MANUAL WLROL EN E 8881184 WLROL ENj WL ROL 160 1200...
Page 62: ...Page 62...
Page 65: ...Page 65 13 Declaration of Conformity...
Page 66: ...Page 66...
Page 72: ...Page 72 This page intentionally left blank...
Page 73: ...SPA Page 73 This page intentionally left blank...