Page 18
5. Mechanical installation
Risk of foot injuries!
Prevent unit from dropping during installation!
Helping hand of a second person is advisa-
ble.
Risk of electrical shock!
Hazardous electrical voltage.
During installation, the unit must be discon-
nected from power supply and secured
against being switched on again. The
absence of voltage must be ensured by a
measurement.
5.1 Environment Parameters to
be met and mounting recom-
mendations
When selecting the installation site for the
reverse osmosis unit, take the following into
account:
•
The minimum clearances indicated in
the fitting measures section must be
observed in order to ensure adequate
unit ventilation and allow for unob-
structed access in case of maintenance
•
Protection class IP20
•
By design, HygroMatik reverse osmosis
units are not qualified for outdoor installa-
tion (electronic components and water-
bearing parts may be damaged)
•
Ambient temperature must lie between
+5 and +40 °C (+41 and +104 °F) in
order to protect the unit electronics
against damage; frost may damage the
steam cylinder, the solenoid valve and
pump, as well as make hoses burst
•
Avoid environments with excessive
humidity, as it could have negative
effects on the electronics.
•
Installation in closed rooms requires
aeration and, if necessary, temperature
conditioning in order to meet the envi-
ronmental conditions.
•
Make use of existing water connections
for supply and draining.
•
For proper functioning of the level con-
trol, plumb and level installation of the
unit is required.
•
The optimal feed water pressure for the
reverse osmosis units is in the range of
2 bar to 5 bar. If these values are not
reached, the installation of a pressure
reducer or a booster pump is neces-
sary.
•
The on-site waste water connection
must be a free outflow according to DIN
EN 1717.
•
The installation room must be well ven-
tilated and equipped with a floor drain.
Alternatively, a water stop device (e.g.
leakage detector) can be used.
•
Ensure that the system stands firmly.
WARNING
!
WARNING
!
Summary of Contents for WL-ROL 160-1200 Series
Page 1: ...FlexLine Reverse osmosis plants MANUAL WLROL EN E 8881184 WLROL ENj WL ROL 160 1200...
Page 62: ...Page 62...
Page 65: ...Page 65 13 Declaration of Conformity...
Page 66: ...Page 66...
Page 72: ...Page 72 This page intentionally left blank...
Page 73: ...SPA Page 73 This page intentionally left blank...