INS0057 EN 

   DEC 2014 

Rev. E 

 

Directions for Use - 

Aquasnap™ Total & Aquasnap™ Free  

ATP Tests for Water Samples 

Part Number:  Aquasnap Total - AQ-100X (100 tests)

 

 

Aquasnap Free - AQ-100FX (100 tests) 

 

Description / Intended Use: 

Aquasnap

TM

 water testing devices are self-contained ATP tests 

for use with Hygiena luminometers. Tests are used to monitor 
ATP levels in water as a quality indicator in areas like Clean-In-
Place (CIP) systems and rinse water samples. It is also used for 
water treatment applications in healthcare and monitoring 
biomass in cooling towers. Aquasnap

TM 

Total measures both 

ATP contained within living cells and particulate matter 
(microbial ATP) as well as ATP dissolved in water (non-
microbial or dead microbial ATP). Aquasnap

TM

 Free measures 

only dissolved ATP outside of living cells (non-microbial ATP). 
Used together, Aquasnap

TM

 Total and Aquasnap

TM

 Free can be 

an effective quality monitoring system. The difference between 
Total and Free test results represents ATP from living 
organisms (also referred to as biomass).

 

Total ATP= Free ATP + Microbial ATP, therefore, 
Microbial ATP = Total ATP 

– Free ATP 

 
For more information about differences between Aquasnap

TM

 

Total and Aquasnap

TM

 Free, visit 

www.hygiena.com

 

 
Directions:  

Instructional Video:

 www.youtube.com/HygienaTV 

 

1. 

Allow Aquasnap

TM

 test devices to equilibrate to room 

temperature (21-25°C) before use. Forcefully flick device in 
a downward motion to shake liquid extractant from 
collection dipper to bottom of tube. Shaking liquid 
extractant to bottom of tube assists with accurate 
extraction of ATP and helps collect a more consistent 
sample. Holding swab tube firmly, twist and pull collection 
device out of tube. Submerge sample collection dipper in 
water sample for 1-2 seconds.  
 

2. 

Lift sample collection device up vertically and reinsert in 
test tube. Gently shake device for 1-2 seconds to release 
water sample from collection tip and to mix sample with 
extractant at bottom of test tube. 
 

3. 

Activate

 

Aquasnap

TM

 by holding tube firmly and using 

thumb and forefinger to break 

Snap Valve™ by bending 

bulb forward and backward. Squeeze bulb twice, expelling 
all liquid into tube. 
 

4. 

Shake for 3-5 seconds to mix sample. 
 

5. 

Insert entire Aquasnap

TM

 device into luminometer, close lid 

and press “OK” to initiate measurements.  Device should 
be measured within 15 seconds of activation.  
 

6. 

Hygiena recommends setting RLU thresholds according to 
user’s test application. Higher RLU results indicate higher 
contamination in sample. In clean or treated water 
samples, Free ATP results are similar to Total ATP results. 
In some circumstances, when organic matter is present 
and where microbial contamination is low, Total ATP 
results may appear lower than Free ATP results; this is 
normal and is due to presence of extractant in Aquasnap

TM

 

Total. To determine Pass/Fail settings, visit the resource 
library at 

www.hygiena.com

 or email a technical 

representative at

[email protected]

 

 

Sampling

 

Recommendations: 

 

When possible, split water sample into two aliquots. Dip each 
Aquasnap

TM

 test (Total and Free) in a separate aliquot of same 

water sample. If two aliquots of water sample cannot be made 
and two devices have to be dipped in same sample container, 
use Aquasnap

TM

 Free first, then Aquasnap

TM

 Total.

 

Calibration Control: 

It is advisable to run positive and negative controls according to 
Good Laboratory Practices. Hygiena offers the following control: 

 

Calibration Control Kit (Part # PCD4000) 

 

Safety & Precautions: 

Components of Aquasnap

TM

 do not pose any health risk when 

used correctly.

 

1. 

Do not use Aquasnap

TM

 test if it is already activated. 

2. 

Hold luminometer upright when measuring test devices. 

3. 

Hold Aquasnap

TM

 test upright when activating. 

4. 

Read Aquasnap

TM

 test within 15 seconds of activation. 

5. 

Keep Aquasnap

TM

 test out of direct sunlight. 

 

Storage & Shelf Life: 

Store at 2-8°C. Devices may be stored at room temperature  
(21-25°C) for up to 4 weeks. Aquasnap

TM

 has a shelf life of 15 

months. Check expiration date on label. 
 

Hygiena Liability: 

Hygiena will not be liable to user or others for any loss or damage 
whether direct or indirect, incidental or consequential from use of 
this device. If this product is prov

en to be defective, Hygiena’s 

sole obligation will be to replace product or at its discretion, refund 
the purchase price. Promptly notify Hygiena within 5 days of 
discovery of any suspected defect and return product to Hygiena. 
 

Contact information: 

If more information is required, please visit us at 

www.hygiena.com

 or contact us at: 

Hygiena - Americas 

Phone: 805.388.8007 
Fax: 805.388.5531 
Email

[email protected]

 

Hygiena - International 

Phone: +44 (0)1923 818821 
Fax: +44 (0)1923 818825 
Email

[email protected]

 

Summary of Contents for AQ-100FX

Page 1: ...lid and press OK to initiate measurements Device should be measured within 15 seconds of activation 6 Hygiena recommends setting RLU thresholds according to user s test application Higher RLU results...

Page 2: ...tivo debe medirse dentro de los 15 segundos de la activaci n 6 Hygiena recomienda configurar los l mites de las URL seg n la aplicaci n de la prueba del usuario Los resultados m s altos de las URL ind...

Page 3: ...enaTV 1 AquasnapTM 21 25 C ATP 1 2 2 1 2 3 AquasnapTM Snap ValveTM 4 3 5 5 AquasnapTM OK 15 6 Hygiena RLU RLU Free ATP Total ATP Total ATP Free ATP AquasnapTM Total www hygiena com enquiries hygiena c...

Page 4: ...m HygienaTV 1 panqauqAA 21 25 PTA 1 2 2 1 2 3 panqauqAA SyHp VHlvg 4 3 5 5 panqauqAA KO 15 6 aneiFuq ULR ULR eeFF plA latqT plA latqT plA eeFF plA panqauqAA latqT www hygiena com enquiries hygiena com...

Page 5: ...2 1 2 3 AquasnapTM Snap Valve 4 3 5 5 AquasnapTM OK 15 6 Hygiena Free Total Total Free ATP AquasnapTM Total www hygiena com enquiries hygiena com AquasnapTM Total Free AquasnapTM Free AquasnapTM Tota...

Page 6: ...6 A Hygiena recomenda configurar os limites de RLUs de acordo com a aplica o de teste do usu rio Resultados mais altos de RLUs indicam maior contamina o na amostra Em amostras de gua limpa ou tratada...

Page 7: ...1 2 2 1 2 3 AquasnapTM Snap Valve 2 4 3 5 5 AquasnapTM OK 15 6 Hygiena RLU RLU Free ATP Total ATP Total ATP Free ATP AquasnapTM Total www hygiena com E enquiries hygiena com 2 AquasnapTM Total Free 2...

Page 8: ...fiehlt diese RLU Grenzwerte gem der Testanwendung des Benutzers zu bestimmen H here RLU Ergebnisse zeigen eine h here Kontamination in der Probe an Bei Proben mit sauberem oder behandeltem Wasser hnel...

Page 9: ...s mesures Le dispositif devrait tre mesur apr s 15 secondes d activation 6 Hygiena recommande que les seuils de param trage URL soient conformes aux applications de test de l utilisateur Des r sultats...

Page 10: ...ntro 15 secondi dall attivazione 6 Hygiena raccomanda di impostare i valori di soglia ULR in base agli standard della propria struttura Valori ULR pi elevati sono indicativi di contaminazione del camp...

Reviews: