FAM 15/30/50/70
Starting up the FAM
HYDAC Filter Systems GmbH
en
Page
36 /
68
BEWA FAM15-70Siem 3099574e en 2009-10-08.doc
2008-10-06
Starting up the FAM
NOTICE
Connection IN / OUT closed
The unit will be destroyed
►
Check to be sure that all of the locking fixtures at the inlet / outlet are in
"open" position each time before start-up.
Switching on the FAM
Move main switch to the "ON" position and wait until the "Fault" lamp lights up briefly
(lamp test) and this message: appears on the display.
Ready for operation
--> Press start
Press the
key. Now the system starts with the
starting phase
.
FluidAqua operating
Starting
1. Vacuum pump starts,
2. inlet valve opens and the vacuum chamber is filled.
3. If the fill level 02 is reached, the evacuation pump switches on and the
starting phase ends.
4. Automatic mode is active:
FluidAqua operating
Automatic mode
Now regulate the vacuum pressure with the throttle valve on the right of the vacuum
gauge, depending on the operating viscosity.
Operating viscosity
Pressure setting (absolute pressure)
50 mm
2
/s hydraulic oil
350 mbar / 35 kPa
300 mm
2
/s lubrication oil
500 mbar / 50 kPa
50 mm
2
/s ester oil
400 mbar / 40 kPa
Some oils begin to foam intensively when they enter the vacuum chamber. This can
especially be the case when the water content is very high, but also when the oil is
old or high viscosity media are used. If too much foam is caused, it is drawn in by
the vacuum pump.
In this case, for example, reduce the vacuum from 350mbar to 450mbar.