FAM 15/30/50/70
Troubleshooting
HYDAC Filter Systems GmbH
en
Page
46 /
68
BEWA FAM15-70Siem 3099574e en 2009-10-08.doc
2008-10-06
Error message
Cause
Remedy
Page
pump: Drain the fluid,
flush and fill with fresh
water.
Rotary vane vacuum
pump: Replace oil and
check air-trapping
element, replace if
necessary.
Incorrect phase sequence
Wrong phase sequence
or operation voltage too
low
Exchange two of the
three phases, check the
supply voltage
Dry running E-pump
Vacuum chamber is
being filled too slowly,
flow into reactor is
smaller than flow rate of
the evacuation pump.
Check suction hose / oil
level in tank. Clean
suction filter, if installed.
Check vacuum pump
and clean if necessary
Outlet (OUT) clogged
No operating fluid is
being pumped out of the
FAM.
Open shut-off devices in
suction line.
Inlet (IN) clogged
No fluid is being
pumped into the FAM
Open shut-off devices in
suction line.