7 | Operation
HYDAC FILTER SYSTEMS GMBH
96 / 148
BeWa FCU1310 3371346g en-us lq
7.7.2 Extracting / copying measurement reports via USB
interface
Saved measurement reports can be copied to the USB mem-
ory stick supplied with the unit. The data in the internal mem-
ory is not deleted while doing so.
Afterwards, delete the measurement reports explicitly in the in-
ternal measured data memory. Details regarding this can be
found in chapter xx.
NOTICE
No recording of measurement reports during copying
Data loss
u
Do not copy the measurement reports from the internal
memory during an ongoing measurement.
NOTICE
Insufficient memory space on the USB memory stick
Data loss
u
Use only a USB memory stick with ≥ 10 MB free memory
space.
Format the USB memory stick prior to its first
use. To do that, insert it into a free USB port on
your PC. Then change to the file manager (e.g.
Explorer) and format the stick in FAT32 format.
You will find details of this in the documentation
of your operating system.
Compatibility with other USB memory sticks
cannot be guaranteed as the FCU communi-
cates directly with the microprocessor. This
Summary of Contents for FCU 1310
Page 148: ......