General Safety Information
CTM-EF 1000 series
en(us)
Page 12 / 52
BeWa CTM-EF 3509157c en-us 2018-10-18.docx
2018-10-18
Proper/designated use
Use the unit only for the application described in the following.
The CTM-EF is an extraction unit designed to analyze the surface
cleanliness of components by taking wet samples. Quality standards can be
checked and documented by means of the determination of type, size and
quantity of the contamination, and the necessary optimization measures can
be implemented.
Any use extending beyond or deviating therefrom shall not be considered
intended use. HYDAC FILTER SYSTEMS GMBH will assume no liability for
any damage resulting from such use.
Proper or designated use of the product extends to the following:
•
Maintaining adherence to all the instructions contained herein
•
Performing requisite inspection and maintenance work
•
Operation with a permissible test liquid only
NOTICE
Impermissible test liquid
The module will be damaged.
►
Use only test liquids which are compatible with the materials and the
sealing materials they come in contact with.
►
Use G60 Spezial as the test liquid.
►
Do not use completely demineralized water as a test liquid.
Using water containing tensides (permissible pH values: 6 to 10) as a
test liquid is permitted.